Clinical Trials Directory

Trials / Active Not Recruiting

Active Not RecruitingNCT06087718

Feasibility of the Maastro Applicator in Rectal Cancer

Introduction of the Maastro Applicator for Endoluminal Boosting in Rectal Cancer: a Pilot Study

Status
Active Not Recruiting
Phase
N/A
Study type
Interventional
Enrollment
10 (estimated)
Sponsor
Maastricht Radiation Oncology · Academic / Other
Sex
All
Age
18 Years
Healthy volunteers
Not accepted

Summary

The goal of this interventional pilot trial is to confirm that Maastro endoluminal HDR ( High Dose Radiation) contact brachytherapy boosting is feasible and may increase the chance of functional organ sparing of the rectum in patients with rectal cancer. Participants will be treated with chemoradiotherapy and an endoluminal boost with the Maastro applicator.

Detailed description

The goal of this clinical trial is to confirm that Maastro endoluminal HDR contact brachytherapy boosting is feasible and may increase the chance of functional organ sparing of the rectum in patients with rectal cancer. If at least 7 out of 10 planned Maastro applicator treatment series (3 fractions per series) can be conducted successfully from a procedural point of view the treatment will be considered feasible. The study intervention will be similar to the study treatment of arm B of the OPERA trial. Opposed to the treatment in arm B of the OPERA trial, the endoluminal boost will be given using HDR brachytherapy with the Maastro applicator instead of a CXRT (Contact X-ray Radiotherapy) device. The dose profile of the Maastro applicator is similar to the dose profile of CXRT device. As in the OPERA trial patients will be stratified based on tumor size. As the diameter of the treatment field of the largest Maastro applicator (there are two sizes) equals 2.5 cm we will stratify for tumor diameter \< 2.5 cm v ≥ 2.5 cm. In the OPERA trial patients were stratified for a tumor diameter of \< 3.0 cm v ≥ 3.0 cm as currently the largest applicator diameter for the CXRT device is 3.0 cm (currently available applicators: 2.0, 2.5 and 3.0 cm). The endoluminal boost will consist of 3 fractions with a dose equivalent to 30 Gy per fraction prescribed at the surface of the applicator. The 3 boost fractions will be delivered over a 4-week time period (week 1-2-4). Conform OPERA protocol, patients with a tumor size \< 2.5 cm will receive an upfront endoluminal HDR contact boost followed by concurrent chemoradiotherapy (25x1.8 Gy combined with capecitabine 825 mg/m2 bd on radiotherapy days). Patients with a tumor size ≥ 2.5 cm will first undergo concurrent chemoradiotherapy (25x1.8 Gy combined with capecitabine 825 mg/m2 bd on radiotherapy days) to first shrink the tumor and will receive the endoluminal HDR contact boost afterwards in order to eventually fit the tumor surface in the surface of the Maastro applicator.

Conditions

Interventions

TypeNameDescription
DEVICEMaastro applicatorThe dose profile of the Maastro applicator is similar to the dose profile of a CXRT device. Patients will be stratified based on tumor size. The cut-off is set at the maximum diameter of the treatment surface of the Maastro applicator. The boost consists of 3 fractions with a dose equivalent to 30 Gy per fraction prescribed at the surface of the applicator. The 3 boost fractions will be delivered over a 4-week time period. Patients with a tumor size \< 2.5 cm will receive an upfront boost followed by concurrent chemoradiotherapy (25x1.8 Gy combined with capecitabine 825 mg/m2 bd on radiotherapy days). Patients with a tumor size ≥ 2.5cm will first undergo concurrent chemoradiotherapy (25x1.8 Gy combined with capecitabine 825 mg/m2 bd on radiotherapy days) and will afterwards receive the boost.

Timeline

Start date
2024-08-27
Primary completion
2026-04-01
Completion
2031-04-01
First posted
2023-10-18
Last updated
2025-08-14

Locations

1 site across 1 country: Netherlands

Source: ClinicalTrials.gov record NCT06087718. Inclusion in this directory is not an endorsement.