Clinical Trials Directory

Trials / Completed

CompletedNCT04692571

EMG-Based Hand-Wrist Control: Study B Mirrored

EMG-Based Hand-Wrist Control: Study B: Comparison of Contralateral (Mirrored) EMG-

Status
Completed
Phase
EARLY_Phase 1
Study type
Interventional
Enrollment
20 (actual)
Sponsor
Liberating Technologies, Inc. · Industry
Sex
All
Age
18 Years
Healthy volunteers
Accepted

Summary

Study assessing four-channel prosthesis controller, that compares contralateral (mirrored) EMG-force training to ipsilateral EMG-target training with both limb-absent and able-bodied subjects

Detailed description

System identification models relating forearm electromyogram (EMG) signals to phantom wrist radial-ulnar deviation force, pronation-supination moment and/or hand open-close force (EMG-force) are hampered by lack of supervised force/moment output signals in limb-absent subjects. In able-bodied and unilateral transradial limb-absent subjects, we studied three alternative supervised output sources in one degree of freedom (DoF) and 2-DoF target tracking tasks: (1) bilateral tracking with force feedback from the contralateral side (non-dominant for able-bodied/ sound for limb-absent subjects) with the contralateral force as the output, (2) bilateral tracking with force feedback from the contralateral side with the target as the output, and (3) dominant/limb-absent side unilateral target tracking without feedback and the target used as the output.

Conditions

Interventions

TypeNameDescription
DEVICEPSICON Measurement ApparatusTest apparatus acquires 16 sEMG channels, measure four DoFs of force/moment at the hand-wrist \[hand open-close (Opn-Cls); wrist extension-flexion (Ext-Flx), radial-ulnar deviation (Rad-Uln) and pronation-supination (Pro-Sup)\]. sEMG system used 16 custom encased bipolar electrodes. A commercial LTI EMG amplifier (BE328) conditioned each EMG signal before A/D conversion on a PC. A thermoplastic hand cuff secured the wrist to a six-axis load cell (AMTI, model MC3A-100) for force/moment measurement. A separate hand-grasp one-axis load cell (Omega Engineering Inc., model LCR-150) was Velcro-secured between the fingers and the thumb, to measure power grip forces. The PC acquired and stored the sEMG and load cell data, and commanded a triangular screen target. The subject controlled a second triangular cursor. Left-right movement of the cursor was controlled by wrist Ext-Flx, up-down movement by Rad-Uln deviation, cursor rotation by Pro-Sup and cursor size by hand Opn-Cls forces.

Timeline

Start date
2017-06-29
Primary completion
2020-08-24
Completion
2020-08-24
First posted
2021-01-05
Last updated
2021-01-05

Locations

2 sites across 1 country: United States

Source: ClinicalTrials.gov record NCT04692571. Inclusion in this directory is not an endorsement.