Clinical Trials Directory

Trials / Completed

CompletedNCT02810080

FiO2 Closed Loop Control Ventilation

FiO2 Closed Loop Control Using the ZOLL 731 Series Ventilator

Status
Completed
Phase
Study type
Interventional
Enrollment
210 (actual)
Sponsor
Zoll Medical Corporation · Industry
Sex
All
Age
18 Years – 65 Years
Healthy volunteers
Not accepted

Summary

A randomized, multi-center trial involving critically-ill patients comparing conventional, manual control of inspired oxygen (FiO2) within a target range, with a computer-controlled, closed-loop FiO2 delivery using a continuous ventilator.

Detailed description

The trial will use a proportional-integral-derivative (PID) type controller for automatic adjustment of FiO2 to maintain the SpO2 within a target range, based on continuous use of pulse oximetry. Intubated subjects will be randomized to either manual control of FiO2 or closed-loop control of FiO2, during a 12-hour period. The PCLC is designed to respond to acute hypoxemia (SpO2 \<88%) within seconds while preventing hypoxemia and minimizing exposure to excessive inspired oxygen levels.

Conditions

Interventions

TypeNameDescription
DEVICEClosed Loop Control VentilationClosed loop control ventilation using ZOLL 731 ventilator with physiologic closed-loop control software
DEVICEManual Control VentilationManual control ventilation using ZOLL 731 ventilator

Timeline

Start date
2016-12-01
Primary completion
2020-12-01
Completion
2020-12-01
First posted
2016-06-22
Last updated
2024-08-15
Results posted
2024-08-15

Locations

1 site across 1 country: United States

Source: ClinicalTrials.gov record NCT02810080. Inclusion in this directory is not an endorsement.