Clinical Trials Directory

Trials / Completed

CompletedNCT02247284

Reproducibility of RNFL and ONH Measurements in OAG With Heidelberg Spectralis SD-OCT Glaucoma Module Premium Edition

Study to Investigate Reproducibility of the New Heidelberg Spectralis SD-OCT Glaucoma Module Premium Edition (Software Version 6.0) in Comparison to Reproducibility of RNFL Scans.

Status
Completed
Phase
N/A
Study type
Interventional
Enrollment
50 (actual)
Sponsor
University of Zurich · Academic / Other
Sex
All
Age
18 Years
Healthy volunteers
Not accepted

Summary

The purpose of this study is to assess and compare the repeatability of SD-OCT (Spectralis™ OCT, Version 6.0, Heidelberg Engineering, Germany) RNFL thickness measurements with a new software (glaucoma premium module) with the well established peripapillary RNFL thickness measurements in patients with primary open angle glaucoma.

Detailed description

The purpose of this study is to assess and compare the repeatability of SD-OCT (Spectralis™ OCT, Version 6.0, Heidelberg Engineering, Germany) RNFL thickness measurements with the new software (method B) and with the previous method of measurement (method A) in patients with glaucoma. It is a prospective, not blinded, not randomised, monocentric study. Inclusion of 50 patients with a diagnosis of primary open angle glaucoma (POAG), that will be recruited from the ophthalmology outpatient department of the UniversityHospital Zurich, Switzerland. All patients will undergo a full ophthalmic examination including measurement of refraction, best corrected visual acuity, examination of anterior and posterior compartment and measurement of the intraocular pressure. On the same day they will have 3 OCT examinations with a new method of measurement (method B).On the same day the first 10 patients will undergo also 3 OCT examinations with the previous method (method A) and they will get an other 3 OCT examinations with method B by a second observer. For this 10 patients the next consultation (3 months +/- 1 month) the 3 OCT examinations with method B will be repeated by the first observer. Primary objectives are coefficient of variations (COV) of method A and B; Intra-class-correlations (ICC) of method A and B.

Conditions

Interventions

TypeNameDescription
DEVICERNFL and BMO-MRW SD-OCTMasurement of peripapillary RNFL measurements with Heidelberg Spectralis SD-OCT (Heidelberg Engeneering GmbH, Heidelberg, Germany) old protocoll and in addition with the new Glaucoma Premium Module (software version 6.0) tocoll, which in addition to RNFL measurements include measurement of Bruch's Membrane Opening-based minimum rim width (BMO-MRW).

Timeline

Start date
2015-03-01
Primary completion
2019-12-31
Completion
2019-12-31
First posted
2014-09-25
Last updated
2020-03-25

Locations

1 site across 1 country: Switzerland

Source: ClinicalTrials.gov record NCT02247284. Inclusion in this directory is not an endorsement.