Clinical Trials Directory

Trials / Recruiting

RecruitingNCT07200674

Interest of High-speed Centrifugation for Measuring Anti-Xa Activity on Lipemic Plasma

Status
Recruiting
Phase
Study type
Observational
Enrollment
150 (estimated)
Sponsor
University Hospital, Strasbourg, France · Academic / Other
Sex
All
Age
18 Years
Healthy volunteers
Not accepted

Summary

Some hemostasis techniques require spectrophotometric measurements. These are then subject to interference from the sample's lipemia (postprandial, hypertriglyceridemia, etc.), which risks distorting the results; this is estimated to affect 0.5 to 2.5% of samples. This is particularly the case for anti-Xa activity, a critical test used to monitor anticoagulant treatments. To date, the HUS Hematology Laboratory does not have or is aware of any alternative method to overcome this interference, and is unable to perform anti-Xa activity when the sample has excessive lipemia. This study will evaluate, on samples from routine care artificially overloaded with lipids, the effectiveness of a high-speed centrifugation process (10,800g for 10 minutes).

Conditions

Timeline

Start date
2024-03-11
Primary completion
2025-12-11
Completion
2025-12-11
First posted
2025-10-01
Last updated
2025-10-01

Locations

1 site across 1 country: France

Source: ClinicalTrials.gov record NCT07200674. Inclusion in this directory is not an endorsement.