Assessment of Tooth Mobility Using the Osstell Device

Objective Assessment of Tooth Mobility Using the Osstell Device: A Custom-Bonded Apparatus Trial

Status
Completed
Phase
N/A
Study type
Interventional
Enrollment
16 (actual)
Sponsor
Mustafa Kemal University · Academic / Other
Sex
All
Age
18 Years – 45 Years
Healthy volunteers
Not accepted

Summary

In this trial, it is hypothesized that mobility values obtained from natural teeth using the Osstell device-adapted with a specifically designed attachment-will demonstrate a statistically significant correlation with those obtained through established methods such as the Periotest and the Miller index. Accordingly, the aim of this study is to investigate the applicability of the Osstell device for assessing the mobility of natural teeth having different mobility grades and to evaluate the correlation between its measurements and those derived from conventional clinical mobility assessment methods.

Detailed description

A total of 16 patients, including 10 males and 6 females, were selected for the study, and mobility measurements were performed on 94 permanent teeth. All measurements were performed twice on each tooth by the same calibrated investigator to ensure consistency and eliminate inter-examiner variability. The repeated measurements were carried out under identical conditions for each assessment method (Miller mobility grading, Periotest M, and Osstell Beacon). To evaluate the intra-examiner reliability, the agreement of the two consecutive measurements was assessed using appropriate statistical methods. High intra-examiner agreement was considered indicative of the measurement reliability and reproducibility within the study. To assess tooth stability, resonance frequency analysis was performed using the Osstell Beacon device (Osstell, AB, Gothenburg, Sweden). A custom-fabricated titanium bracket was used to accommodate the smartPeg and ensure standardized placement on each tooth. The bracket-like apparatus was carefully aligned with the long axis of the tooth and positioned apically. Bracket specifications * Raw material: Medical-grade titanium, precision-machined using Computer Numerical Control technology. * Design: Cylindrical base with a centrally located, threaded slot for smartPeg insertion SmartPeg thread slot * Internally threaded to enable clockwise insertion of the smartPeg (Figures 3b and 3c) * Thread type: micro-thread, * Tapered, centred aperture * Diameter: approximately 1.2-1.5 mm Tooth contact surface * Flat base designed to conform to the buccal surface of the tooth * Smooth texture with micro-mechanical retention properties * Suitable for adhesive fixation or direct contact application The apparatus was placed on the buccal surface of the respective tooth using flowable composite. Care was taken to ensure that the slot of the bracket designated for a smartPeg placement was oriented toward the root. Subsequently, the smartPeg was positioned in alignment with the root and torqued into place. After the smartPeg installation, two measurements were taken from different directions. The mean of the two readings was recorded as the final RFA value for each tooth. The Osstell Beacon was operated at a 2-5 mm distance, held perpendicular (90°) to the smartPeg without making contact. In this approach, the unit of measurement is the Implant Stability Quotient (ISQ), which is derived from resonance frequency analysis and ranges from 0 to 100, with higher values indicating greater stiffness at the implant-bone interface.The normality of the data distribution was tested using the Shapiro-Wilk test. The relationships between numerical variables were evaluated using Spearman's correlation coefficient. To assess the agreement between mobility classes and the Periotest and Osstell groups, Cohen's Kappa statistic was employed. The Kappa coefficient was interpreted as follows: values \<0.00 indicated "poor agreement," 0.00-0.20 "slight agreement," 0.21-0.40 "fair agreement," 0.41-0.60 "moderate agreement," 0.61-0.80 "substantial agreement," and 0.81-1.00 "almost perfect agreement." The reliability of stability scores obtained from the Periotest and Osstell devices was evaluated using the Intraclass Correlation Coefficient (ICC). To determine inter-device agreement, ICC values were calculated based on a two-way mixed-effects model using the single measurement (average measure) approach. ICC values ≥0.75 were considered to indicate satisfactory reliability. All statistical analyses were performed using SPSS (Statistical Package for the Social Sciences), with a significance level set at 5% (p\<0.05)

Conditions

Interventions

TypeNameDescription
DEVICEMiller mobility measurementTooth mobility at different levels of mobility was assessed using Miller, Periotest and Osstell, and was evaluated both subjectively and objectively.

Timeline

Start date
2022-06-01
Primary completion
2022-12-31
Completion
2023-12-31
First posted
2025-09-23
Last updated
2025-09-23

Locations

1 site across 1 country: Turkey (Türkiye)