Closed Loop Oxygen Control in INtubated Critically Ill Patients
Closed Loop Oxygen Control in INtubated Critically Ill Patients - CLOC-IN
- Status
- Recruiting
- Phase
- N/A
- Study type
- Interventional
- Enrollment
- 46 (estimated)
- Sponsor
- Parc Taulí Hospital Universitari · Academic / Other
- Sex
- All
- Age
- 18 Years
- Healthy volunteers
- Not accepted
Summary
The deleterious effects of alveolar hyperoxia and (severe) hyperoxaemia are well described. Achieving safe and efficient oxygenation may be challenging. The hypothesis of the present study is that closed-loop oxygen control in intubated and mechanically ventilated critically ill patients improves oxygen administration compared with standard manual oxygen titrations. A single-blind, randomised crossover clinical trial assessing the efficacy and safety of the use of a closed-loop oxygen control versus manual oxygen titration in patients receiving mechanical ventilation has been designed. Patients will be randomised to receive first either closed-loop oxygen control (CLOC) or manual oxygen titration (MOT). The percentage of time spent in optimal and sub-optimal SpO2 (oxygen saturation by pulse oximetry) ranges will be calculated for each period.
Detailed description
The deleterious effects of alveolar hyperoxia and (severe) hyperoxaemia are well described. A safe and efficient oxygenation should be achieved. However, from a practical point of view, this may require frequent manual adjustments of the inspired oxygen, which might be unfeasible, especially in high-demand periods. Several automatic oxygen systems have been tested in non-intubated patients. The hypothesis of the present study is that closed-loop oxygen control in intubated and mechanically ventilated critically ill patients improves oxygen administration compared with standard manual oxygen titrations. Thus, the objective of the study is to compare the percentage of time spent in a predefined optimal SpO2 range between closed-loop oxygen control and manual oxygen titration. A single-blind, randomised crossover clinical trial assessing the efficacy and safety of the use of a closed-loop oxygen control versus manual oxygen titration in patients receiving mechanical ventilation has been designed. Once included, patients will be randomised to receive first either closed-loop oxygen control (CLOC) or manual oxygen titration (MOT). Every period will last for 4h. Clinical and respiratory variables will be recorded. A memory box will be connected to the patient's ventilator. This device will record ventilator parameters, including FiO2 and SpO2. To assess the outcome, the percentage of time spent in optimal and sub-optimal SpO2 ranges will be calculated for each period.
Conditions
Interventions
| Type | Name | Description |
|---|---|---|
| DEVICE | closed-loop oxygen control (CLOC) | HAMILTON-C6 (Hamilton Medical, Bonaduz, Switzerland) ventilators will be used. It allows for O2 Assist software, that will be installed on all C6 ventilators before the start of this study. Automatic Oxygen management sets the Oxygen control according to the patient's measured SpO2 and the operator-set SpO2 target range. The Oxygenation controller will be activated for 4h and adjusts the Oxygen according to patient's measured SpO2. |
| DEVICE | 2. manual oxygen titration (MOT) | O2 will be manually set according to SpO2 target during a 4h period. |