Multiplo Tp/HIV Self-Test
A Study to Evaluate the Accuracy, Usability and Readability of the Multiplo® TP/HIV Antibody Self-Test Performed by Observed Intended Users in Canada and the Impact of Peer-led HIV and Syphilis Care Among People Experiencing Homelessness in Toronto
- Status
- Recruiting
- Phase
- N/A
- Study type
- Interventional
- Enrollment
- 900 (estimated)
- Sponsor
- Unity Health Toronto · Academic / Other
- Sex
- All
- Age
- 18 Years
- Healthy volunteers
- Accepted
Summary
To help reach the undiagnosed living with HIV and/or syphilis in Canada, self-tests for HIV and Syphilis may have substantial utility for increased identification of infected individuals through their relative ease of use and portability, as well as their ability to deliver rapid, actionable results while the care provider still has access to the patient. MedMira Laboratories Inc. (Halifax, Nova Scotia, Canada) has developed a point-of-care (POC) test to detect HIV and Syphilis antibodies in fingerstick blood samples that is under final review by Health Canada for use by trained Healthcare professionals. A self-test version of this test, with simplified instructions for use has been developed for investigational studies. The goal of the following study sponsored by REACH Nexus is to provide evidence that untrained lay persons / intended users can perform the Multiplo Tp/HIV Self-Test without any increased risk of obtaining erroneous results.
Detailed description
This is a cross-sectional study that employs both observational and interview-based methods. All participants will read the instructions for use (IFU) and conduct the Multiplo TP/HIV Self-Test. They will not be provided any training prior to use. A trained Observer will observe and document the use of the test and independently read the lay user self-test result. 400 participants will also be provided with ten (10) mock devices with different results and asked to interpret those results (i.e. strong positive, weak positive, negative, invalid with no control line and no test dots, and invalid with no control line and positive test dots). The participants will then complete a questionnaire to share their opinions on usability and label comprehension of this test. All participants will also have a venous blood sample collected for testing of serum at a central laboratory with a 4th generation HIV EIA and standard syphilis serology that will serve as the Comparator Methods (CM). All self-tests will be confirmed using laboratory-standard HIV and syphilis testing algorithms. The sensitivity and specificity of the self-test results will be calculated relative to the "clinical truth" of the participants' HIV and syphilis status determined by the comparator methods. All enrolled Participants will voluntarily provide medical history and the specimens according to Section 4, Study Procedures at Visit 1 (Day 1), for the testing with the Multiplo TP/HIV Self-Test. Persons with reactive syphilis self-tests may be counselled for immediate treatment. All Participants will have a laboratory test done for performance comparison and will be instructed to return to the clinic for a follow-up visit (Visit 2), two (2) weeks later to obtain their laboratory test results. Participants may also have a separate Point-of-Care (POC) test for HIV and/or syphilis conducted on-site as part of the clinic's standard of care testing, outside of the protocol. * Subject recruitment * Registration and confirmation of eligibility (enrolment questionnaire) * Administering informed consent * Review of inclusion/exclusion criteria and enrollment * Venous blood specimen collection for laboratory testing * Participant brought to a private room with site observer * Observer instructs participant on study process (not testing procedure) * Participant given one Multiplo Tp/HIV Self-Test package * Participant performs the self-test and interprets the results * Observer then interprets the participant's self test results * Observer interviews participant with self-test questionnaire * Participant given the option of participating in the Mock Result Interpretation exercise * Participant returns to the site for the laboratory results in approximately 2 weeks MOCK RESULTS INTERPRETATION A minimum of 400 participants across all study sites will complete device interpretation of contrived (mock) results from prepared devices. Devices will include contrived reactive, weakly reactive, negative, and 2 types of invalid results with their reactivity blinded to the participant.
Conditions
Interventions
| Type | Name | Description |
|---|---|---|
| DEVICE | Multiplo TP/HIV Antibody Test | The Multiplo TP/HIV Antibody Test (MedMira Laboratories Inc., Halifax, NS), \["Multiplo TP/HIV Test"\] is a single use, rapid, vertical-flow in vitro qualitative immunoassay for the detection of antibodies to Human Immunodeficiency Virus (HIV) Type 1/Type 2 and Treponema pallidum (Syphilis) in human fingerstick blood. The Multiplo® TP/HIV Test is intended as an aid in the diagnosis of HIV1/2 and Syphilis infections in patients with signs and symptoms of HIV and syphilis. The test is intended for use by trained personnel in medical facilities, clinical laboratories, emergency care situations, and physicians' offices as an in vitro diagnostic device capable of providing results in less than five minutes. The Multiplo TP/HIV Test has not yet sought approval for self-testing. All required pre and post-test counselling guidelines must be followed in each setting in which the Multiplo TP/HIV Test is used. Results are read visually. |
Timeline
- Start date
- 2025-06-12
- Primary completion
- 2026-04-01
- Completion
- 2026-12-10
- First posted
- 2024-12-04
- Last updated
- 2026-01-27
Locations
3 sites across 1 country: Canada
Source: ClinicalTrials.gov record NCT06716450. Inclusion in this directory is not an endorsement.