Clinical Trials Directory

Trials / Completed

CompletedNCT06447896

Clinical Validation of a Cardiovascular Autonomic Nervous Function Multi-Parameter Evaluation System

A Open-label, Randomised, Paired Design, Multicenter Study of a Cardiovascular Autonomic Nervous Function Multi-Parameter Evaluation System With Intelligent Human-Machine Interaction for Assessing Cardiovascular Autonomic Neuropathy

Status
Completed
Phase
N/A
Study type
Interventional
Enrollment
235 (actual)
Sponsor
REEM (Shenzhen) Healthcare Co., Ltd · Industry
Sex
All
Age
18 Years
Healthy volunteers
Accepted

Summary

This is a open-label, randomized, paired design, multicenter study of the Cardiovascular Autonomic Nervous Function Multi-Parameter Evaluation System (R6000) for evaluating cardiovascular autonomic neuropathy. R6000 is novel intellectual diagnostic device for cardiovascular autonomic reflex tests and heart rate variability analysis. A total of 85 subjects will be enrolled in trial Ⅰ and 150 subjects will be enrolled in trial Ⅱ. In trial Ⅰ, subjects will be randomized to two groups, one of which will receive test by the tested device first and the other one will be tested by conventional blood pressure kit. The primary objective of this clinical trial is to evaluate the effectiveness of the Cardiovascular Autonomic Nervous Function Multi-Parameter Evaluation System. The secondary objective is to evaluate the safety of the cardiovascular autonomic testing system and stability.

Detailed description

Trial I Tested device: Cardiovascular Autonomic Nervous Function Multi-Parameter Evaluation System (R6000). Control: Conventional blood pressure kit. Goal: To assess the consistency between the tested device and the conventional method in measuring blood pressure. Trial IIa Tested device: Cardiovascular Autonomic Nervous Function Multi-Parameter Evaluation System (R6000). Control: Ambulatory holter station. Goal: To assess the consistency between the tested device and the conventional devices in measuring heart rate variability (HRV). Trial IIb Tested device: Cardiovascular Autonomic Nervous Function Multi-Parameter Evaluation System (R6000). Control: Manual method with digital ECG machine, stethoscope and sphygmomanometer. Goal: To assess the consistency between the tested device and the manual method in determining cardiovascular autonomic neuropathy (CAN).

Conditions

Interventions

TypeNameDescription
DEVICECardiovascular Autonomic Nervous Function Multi-Parameter Evaluation System (Model: R6000)Cardiovascular autonomic reflex tests (CARTs) and heart rate variability analysis (HRV) are conducted with Cardiovascular Autonomic Nervous Function Multi-Parameter Evaluation System (Model: R6000). CARTs include five standardized tests, which are heart rate response to lying-to-standing, Valsalva maneuver, deep breathing, and blood pressure response to lying-to-standing and sustained isometric handgrip. These series of movements act as "standardized stimuli" to trigger hemodynamic changes in the body, altering vagal and sympathetic activity through mechanisms such as the baroreflex, thereby producing changes in heart rate and blood pressure. The sensor-monitoring interactive design of the device ensures the standardization of these maneuvers.
DIAGNOSTIC_TESTConventional method of CARTsConventional method are conduced using blood pressure kit with manual sphygmomanometer and ECG machine. Conventional method and the investigated device are used simultaneously to record electrocardiograms. Conventional blood pressure taking are carried out in sequence with the investigated device. CARTs results from the conventional method are used as control.
DEVICEAmbulatory holter station Model i12 plusModel i12 plus (manufactured by RENCARE) is an ambulatory holter station, which was approved by the Center for Medical Device Evaluation (CMDE), China National Medical Products Association (NMPA). It can analyze electrocardiogram and generate HRV indices, which include SDNN, RMSSD, NN50, pNN50, triangular index, average RR, deceleration capacity (DC), deceleration runs (DRs), VLF, LF, HF, LF/HF, and total power. Model i12 plus is used to analyzed the same ECG signals as the investigated device.

Timeline

Start date
2011-03-11
Primary completion
2022-08-12
Completion
2022-08-12
First posted
2024-06-07
Last updated
2024-06-10

Source: ClinicalTrials.gov record NCT06447896. Inclusion in this directory is not an endorsement.