Clinical Trials Directory

Trials / Completed

CompletedNCT05779397

Clinical Performance of Masimo Rad-GT

Status
Completed
Phase
N/A
Study type
Interventional
Enrollment
68 (actual)
Sponsor
Masimo Corporation · Industry
Sex
All
Age
18 Years – 80 Years
Healthy volunteers
Accepted

Summary

The study is designed to validate the clinical performance of Masimo Rad-GT in the accuracy of temperature measurements.

Detailed description

Individual sites are only representative of a subset of the overall population, per ISO 80601-2-56:2017(Medical electrical equipment - Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement), and therefore clinical bias and limits of agreement should not be evaluated for the subsets. Publication of results should include repeatability from individual sites and the combined clinical bias and limits of agreement (NCT05779397, NCT05787782, and NCT05674344).

Conditions

Interventions

TypeNameDescription
DEVICEMasimo RAD-GTNon-contact infrared thermometry device

Timeline

Start date
2023-02-20
Primary completion
2023-02-23
Completion
2023-02-23
First posted
2023-03-22
Last updated
2023-07-25
Results posted
2023-07-25

Locations

1 site across 1 country: United States

Regulatory

Source: ClinicalTrials.gov record NCT05779397. Inclusion in this directory is not an endorsement.