Trials / Completed
CompletedNCT05584176
COVID-19 MP Biomedicals SARS-CoV-2 Ag OTC: Clinical Evaluation
Over the Counter Rapid Antigen Test for Detection of SARS-CoV-2 Virus: Clinical Evaluation
- Status
- Completed
- Phase
- N/A
- Study type
- Interventional
- Enrollment
- 332 (actual)
- Sponsor
- MP Biomedicals, LLC · Industry
- Sex
- All
- Age
- 2 Years
- Healthy volunteers
- Accepted
Summary
SARS-CoV-2 rapid antigen over the counter clinical performance evaluation
Detailed description
Coronavirus disease (COVID-19) is a disease caused by a newly discovered coronavirus, severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2). The SARS-CoV-2 is a β-coronavirus, which is enveloped non-segmented positive-sense RNA virus 2. It is spread by human-to-human transmission via droplets or direct contact, and infection has been estimated to have mean incubation period of 6.4 days and a basic reproduction number of 2.24-3.58. Among patients with pneumonia caused by SARS-CoV-2, fever was the most common symptom, followed by cough. On 11 March 2020, the COVID-19 outbreak was characterized as a pandemic by the WHO. Since then, over 70 million people worldwide have been infected with the virus with over 1.5 million deaths attributed to the virus. Laboratory testing for SARS-CoV-2 is currently being performed to determine if an individual has active infection via detection of viral RNA or if an individual has an immune response to the virus from a previous infection via detection of antibodies. Specimen collection is a crucial first step in the evaluation of an individual's SARS-CoV-2 infection status. The goal of this project is to evaluate a rapid SARS-CoV-2 antigen test for over the counter (OTC) use. Study subjects under EDP supervision, either in-person or via video conference, will collect and test anterior nasal swab samples and a Study Representative will collect and ship a second nasopharyngeal swab sample for comparator PCR testing. The Rapid SARS-CoV-2 Antigen Test collection and testing methodology is viewed as a convenient and inexpensive method to test clinical specimens for SARS-CoV-2 and OTC access will improve the availability of COVD-19 testing. The COVID-19 Antigen Rapid Test Device detects SARS-CoV-2 viral antigens through visual interpretation of color development. Anti-SARS-CoV-2 antibodies are immobilized on the test region of the nitrocellulose membrane. Anti-SARS-CoV-2 antibodies conjugated to colored particles are immobilized on the conjugated pad. A sample is added to the extraction buffer which is optimized to release the SARS-CoV-2 antigens from specimen. During testing, the extracted antigens bind to anti-SARS-CoV-2 antibodies conjugated to colored particles. As the specimen migrates along the strip by capillary action and interacts with reagents on the membrane, the complex will be captured by the anti-SARS-CoV-2 antibodies at the test region. Excess colored particles are captured at the internal control zone. The presence of a colored band in the test region indicates a positive result for the SARS-CoV-2 viral antigens, while its absence indicates a negative result. A colored band at the control region serves as a procedural control, indicating that the proper volume of specimen has been added and membrane wicking is working.
Conditions
Interventions
| Type | Name | Description |
|---|---|---|
| DEVICE | iCura COVID-19 Antigen Rapid Home Test | Candidate Ag self-test kit anterior nasal specimen collection and testing will be done under the supervision of qualified site personnel in person. Comparator NP samples shall be collected at least 15 minutes prior to Ag samples with at least 15 minutes between nasal collections including collections that may occur prior to study enrollment. Comparator samples will be collected first by qualified personnel at a study site or physician's office using an FDA authorized nasopharyngeal swab collection kit and SARS-CoV-2 molecular assay. |
| DEVICE | RT-PCR Test | High sensitivity RT-PCR COVID-19 Test |
Timeline
- Start date
- 2021-12-09
- Primary completion
- 2022-01-21
- Completion
- 2022-01-21
- First posted
- 2022-10-18
- Last updated
- 2023-06-01
- Results posted
- 2023-06-01
Locations
4 sites across 1 country: United States
Regulatory
- FDA-regulated device study
Source: ClinicalTrials.gov record NCT05584176. Inclusion in this directory is not an endorsement.