Validation of Vital Signs Recording With VT-Patch Connected Devices in Children

Validation of Vital Signs Recording With VT-Patch® Connected Devices in Children

Status
Recruiting
Phase
Study type
Observational
Enrollment
48 (estimated)
Sponsor
VitalTracer Ltd. · Industry
Sex
All
Age
1 Day – 18 Years
Healthy volunteers
Not accepted

Summary

: Patient readmission to PICU is a major point of concern for physicians because, although being a rare event, it has been strongly associated with more morbidity and death. Patient monitoring once they have left the PICU is drastically modified, and small connected devices could be an option to early identify patients at risk of PICU readmission. Such devices have been manufactured for the recording of vital signs in adults, but they do not exist for children. Besides, studies have recently shown that devices that use optical sensors may be less accurate in individuals with dark skin pigmentation since dark skin tones impose a limitation on optical biometric sensing. The wearable and medical device industry has not sufficiently addressed this issue yet. This is one of the main challenges that need to be addressed in this area to make sure the connected devices using this technology will work with all different skin colours including very dark skin.

Detailed description

Main objective: Assess the accuracy of VT-Patch® to measure oxygen saturation (SpO2) in children with diverse skin colour. Secondary objectives: Assess the accuracy of VT-Patch® to measure 3 additional physiologic parameters (heart rate (HR), respiratory rate (RR), and temperature. Exploratory assessment of blood pressure (BP) measurement, ECG signal, and movement capture. Design: Prospective validation study of VT-Patch® for the recording of 4 physiologic parameters and exploratory assessment of three additional parameters. Methods: The study will be performed in two successive phases: Run-in phase: Testing VT-Patch® on four children to assess the feasibility of the protocol. Main study: validation phase of the devices on 12 children per age group (2 children per skin colour group) to assess its accuracy (VT-Patch® for less than 2 years old). The patients of the run-in phase will be included in the main study if no modification of the protocol is required. 5 physiologic parameters (HR, SpO2, RR, temperature, and ECG) will be prospectively and simultaneously recorded for 8 hours with VT-Patch® and the regular monitoring system implemented in the unit High-resolution database that will be the gold standard for evaluation. Vital signs recorded with the connected device will not be used for the clinical care of patients. The Skin colour of each patient will be assessed by the colorimetric Monk scale and as shown below.

Conditions

Interventions

TypeNameDescription
DEVICENon-invasive monitoringComparison of VitalTracer monitoring devices with invasive and/or non-invasive gold standard reference

Timeline

Start date
2025-02-03
Primary completion
2025-09-01
Completion
2025-12-31
First posted
2022-07-11
Last updated
2025-03-10