Clinical Trials Directory

Trials / Unknown

UnknownNCT05187676

Clinical Evaluation of an Innovative Non-contact Optical Device for Skin Oxygenation Imaging

Study of the Concordance Between Skin Oxygen Saturation Measurements Made by an Innovative Non-contact Optical Imaging Device (IPAM) and the Reference Medical Device (Periflux 6000) in Patients With Skin Vascular Disorders or Chronic Wounds

Status
Unknown
Phase
N/A
Study type
Interventional
Enrollment
70 (estimated)
Sponsor
Central Hospital, Nancy, France · Academic / Other
Sex
All
Age
Healthy volunteers
Not accepted

Summary

According to international consensus, information on the level of skin oxygenation is necessary for the diagnostic management of chronic wounds. The measurement of transcutaneous partial oxygen pressure (or TcPO2, according to the acronym) is measurable with a medical device that uses a Clark electrode. Since the medical device in question is expensive and the measurement time-consuming (approximately 45 minutes per patient), the TcPO2 is only rarely accessible (in hospital medicine as in nursing homes) and is therefore not predominant in the clinical decision. The clinical decision is currently mainly based on the result of the ankle blood pressure measurement, which is readily available. The main objective is to study, from the same skin surfaces (same localization of skin sites and same dimensions of each surface in cm²), in patients with cutaneous vascular disorders or chronic wounds, the relationship between (i) the transcutaneous oxygen saturation values (expressed as a percentage) by the innovative non-contact optical imaging device under study (IPAM, method to be validated) and (ii) the transcutaneous partial oxygen pressure values (TcPO2) (expressed in millimeters of mercury, mmHg) measured by the reference medical device (Périflux6000, gold standard).

Detailed description

The InnovaTICs Dépendance project, funded by the Grand Region Is, aims to develop a medical device meeting the need to measure quickly and non-invasively the skin oxygen saturation. To do this, a device medical was developed by CRAN (Joint Research Unit of the University of Lorraine and CNRS); its industrial transfer was entrusted to the company SD Innovation to produce the VRPC device for "video-reconstruction of chronic wounds". The VRPC medical device makes it possible to measure skin oxygen saturation by an optical (atraumatic) and non-contact (non-invasive) method, more quickly than the current method proposes. The capacity of the VRPC device to measure in vivo variations in skin oxygen saturation was confirmed by carrying out measurements on healthy volunteers. The measured values made it possible to show that the VRPC device was able to detect the difference in skin oxygen saturation between the skin covering a "healthy" finger and the skin covering the same finger but after a tourniquet had been placed at the base of the first phalanx of the finger (data not published).

Conditions

Interventions

TypeNameDescription
DEVICEOxygen saturation measurement using the IPAM deviceSkin oxygen saturation rate will be measured (%) using the non-contact optical device under evaluation called IPAM
DEVICETranscutaneous oxygen partial pressure (TcPO2) using the PeriFlux6000 deviceSkin TcPO2 will be measured (mmHg) using the reference device called PeriFlux6000

Timeline

Start date
2022-01-03
Primary completion
2024-12-31
Completion
2024-12-31
First posted
2022-01-12
Last updated
2022-01-12

Locations

1 site across 1 country: France