Trials / Completed
CompletedNCT05063006
Exercise Tolerance in Patients With Implanted Left Ventricular Assist Device
- Status
- Completed
- Phase
- N/A
- Study type
- Interventional
- Enrollment
- 22 (actual)
- Sponsor
- Karol Wierzbicki · Academic / Other
- Sex
- All
- Age
- 18 Years
- Healthy volunteers
- Not accepted
Summary
To evaluate the concept of dynamic pump speed optimization based on the echocardiographic assessment of aortic valve opening during cardiopulmonary exercise test.
Detailed description
Left ventricular assist devices (LVAD) are becoming a destination therapy in patients with end-stage left ventricular dysfunction. Current generation pumps operate with a fixed rotation speed without the capability of automated speed adjustment. It was shown that acceleration of the pump speed during stress test increases the maximum exercise tolerance. Periodic aortic valve opening (AVO) is used to set up an optimal resting pump speed. The study aimed to evaluate the concept of dynamic pump speed optimization based on the echocardiographic assessment of AVO during the cardiopulmonary exercise test (CPET). Patients with implanted third-generation LVADs with hydrodynamic bearing (HVAD, Medtronic, MN, USA) are prospectively enrolled. Two CPETs are performed after resting speed optimization. The first one with maintained baseline pump speed settings, and the second one with gradually increased speed depending on live echocardiographic imaging. The sequence of tests is random.
Conditions
Interventions
| Type | Name | Description |
|---|---|---|
| OTHER | Cardiopulmonary exercise tests with maintained baseline optimal pump speed settings. | Optimal resting LVAD speed settings is defined as periodic aortic valve opening while maintaining the central position of the intraventricular septum, minimizing mitral regurgitation, and preserving the mean systemic blood pressure above 65 mmHg. The aim is to achieve an aortic valve opening ratio of around 25-33% by changing the pump speed at resting conditions. |
| OTHER | Cardiopulmonary exercise test with gradually increased pump speed depending on live echocardiographic imaging. | During the modified speed exercise, when the aortic valve opening ratio rises above 50%, the pump speed is increased by 100 revolutions per minute (RPM). As the aortic valve remains closed the RPM is lowered by 100. Speed increment is not greater than 100 RPM per 45 seconds to enable an echocardiographic analysis of resulting changes and prevent suction events. There is no determined target speed, neither per time period nor maximal. |
Timeline
- Start date
- 2017-02-15
- Primary completion
- 2020-07-31
- Completion
- 2020-08-04
- First posted
- 2021-09-30
- Last updated
- 2021-10-21
Locations
1 site across 1 country: Poland
Source: ClinicalTrials.gov record NCT05063006. Inclusion in this directory is not an endorsement.