Calibration of a Wrist Cuff Blood Pressure Device, According to the AAMI/ESH/ISO Universal Standard

Calibration of the Cloud DX Pulsewave Health Monitor (PAD-2A) Oscillometric Wrist Cuff Home Blood Pressure Monitor, According to the AAMI/ESH/ISO Universal Standard (ISO 81060-2: AMD_2020) (CCV-4)

Status
Unknown
Phase
Study type
Observational
Enrollment
200 (estimated)
Sponsor
Cardiovascular Research New Brunswick · Academic / Other
Sex
All
Age
19 Years
Healthy volunteers
Accepted

Summary

The objective of this study is to calibrate the blood pressure measurements of the Pulsewave Health Monitor (PAD-2A) device to be within 5±8 mmHg of the average of the dual-observer measurements via an aneroid sphygmomanometer (reference device), as per the methodology described in the ISO 81060-2: AMD\_2020 protocol.

Detailed description

The plan for this study is to calibrate the blood pressure measurements of the PAD-2A device with the dual-observer blood pressure measurements via an aneroid sphygmomanometer (reference device), as per the ISO 81060-2: AMD\_2020 protocol. The data will be collected and analyzed in order to calibrate the PAD-2A device blood pressure measurements to be within 5±8 mmHg of the average of the dual-observer auscultatory blood pressure measurements by adjusting the device algorithm.

Conditions

Interventions

TypeNameDescription
DEVICEPulsewave Health Monitor (PAD-2A) DeviceThis is an observational, comparison study of blood pressure measurements. The objective of this study is to calibrate the blood pressure measurements of the Pulsewave Health Monitor (PAD-2A) device to be within 5±8 mmHg of the average of the dual-observer auscultatory blood pressure measurements via an aneroid sphygmomanometer (reference device), as per the ISO 81060-2: AMD\_2020 universal protocol.

Timeline

Start date
2021-06-15
Primary completion
2023-12-31
Completion
2023-12-31
First posted
2021-05-03
Last updated
2023-02-21

Locations

1 site across 1 country: Canada

Regulatory

Source: ClinicalTrials.gov record NCT04869826. Inclusion in this directory is not an endorsement.

Calibration of a Wrist Cuff Blood Pressure Device, According to the AAMI/ESH/ISO Universal Standard (NCT04869826) · Clinical Trials Directory