Evaluation of Accuracy of CFD-based RuiXin-FFR by Comparing With Pressure-wire-based FFR
Accuracy of RuiXin-FFR by Comparing With Pressure-wire-based FFR in Detecting Hemodynamically Significant Stenosis: A Prospective Multicenter Study
- Status
- Completed
- Phase
- —
- Study type
- Observational
- Enrollment
- 316 (actual)
- Sponsor
- Beijing Anzhen Hospital · Academic / Other
- Sex
- All
- Age
- 18 Years
- Healthy volunteers
- Not accepted
Summary
Coronary Artery Disease (CAD) is the top killer nowadays. Pressure-wire-based Fractional Flow Reserve (FFR) is the gold standard for measuring ischemia in coronary arteries. CFD-based RuiXin-FFR, which is noninvasive, is developed recently. But its accuracy is not verified. This is a multi-center and prospective study to evaluate the sensitivity, specificity, and accuracy of CFD-based RuiXin-FFR compared with wire-based FFR.
Detailed description
The participants are entered with suspended CAD and angiography ischemia between 30%-90% from 6 hospitals in China. Pressure-wire-based FFR is conducted in the standard protocol by each hospital. FFR\>0.8 is chosen as the threshold to indicate non-ischemia. CFD-based RuiXin-FFR is performed by Raysight Medical (Shenzhen, China). CT images are acquired in a standard protocol by each hospital. Images with poor quality are excluded in this trial. Based on these images, RuiXin-FFR reconstructs the 3D model of coronary trees and obtains RuiXin-FFR values by conducting CFD simulations in a blinded fashion. RuiXin-FFR\>0.8 is chosen as the threshold to indicate non-ischemia as well. By comparing the clinical diagnosis outcome of pressure-wire-based FFR and CFD-based RuiXin-FFR, the sensitivity and specificity of the RuiXin-FFR are obtained, which is the primary end point of this trial. The secondary end point includes the measurement of accuracy, NPV, PPV and ROC of the RuiXin-FFR.
Conditions
Interventions
| Type | Name | Description |
|---|---|---|
| DIAGNOSTIC_TEST | Pressure wire fractional flow reserve | Pressure wire is inserted into coronary arteries. The ratio of the average distal pressure and average aorta pressure is FFR. |
| DIAGNOSTIC_TEST | CFD-based RuiXin-FFR | CT-FFR was evaluated in a blinded fashion with a "Coarse-to-Fine Subpixel" algorithm for lumen contour |
Timeline
- Start date
- 2019-03-13
- Primary completion
- 2020-05-14
- Completion
- 2020-08-31
- First posted
- 2021-01-29
- Last updated
- 2021-09-05
Locations
1 site across 1 country: China
Source: ClinicalTrials.gov record NCT04731285. Inclusion in this directory is not an endorsement.