Clinical Trials Directory

Trials / Completed

CompletedNCT04686500

VisionRT-based Deep Inspiration Breath-hold (DIBH) Respiratory Motion Management Strategy, A Pilot Study for Thoracic and Abdominal Tumors

VisionRT-based Deep Inspiration Breath-hold (DIBH) Respiratory Motion Management Strategy, A Pilot Study for Thoracic and Abdominal Tumors Stereotactic Body Radiotherapy

Status
Completed
Phase
N/A
Study type
Interventional
Enrollment
10 (actual)
Sponsor
University of Texas Southwestern Medical Center · Academic / Other
Sex
All
Age
18 Years
Healthy volunteers
Not accepted

Summary

A more recent competing technology for implementing the DIBH technique is real-time surface photogrammetry using the AlignRT system (Vision RT Ltd., London, UK). AlignRT system use non-ionization near infrared light to track patient surface motion. The system has one projector projecting near infrared optical pattern on patient surface. The optical pattern is imaged by optical cameras (two per pod) at \~25 Hz. The user selects a region-of-interest (ROI) on the surface and the software calculates and displays the real-time position in six degrees (3 translations and 3 rotations) in real-time. Once the patient has matched the pre-determined DIBH position (within threshold accuracy), the radiation beam is enabled to be turned on for treatment.

Detailed description

VRT-DIBH has already applied to left breast radiotherapy to spare lung and heart. Compared to ABC-DIBH, VRT-DIBH has several potential advantages: 1. VRT monitors patient (surface) position in addition to DIBH signal, while ABC only check the tidal volume, which can remain the same even if the patient shifts slightly on the couch; 2. VRT is more cost effective, as patient tubing needs to be replaced daily for ABC; 3. VRT potentially has advantage on patient compliance, as with ABC breath is forced impeded while with VRT breath-hold is voluntary. In this study, we renovate the established DIBH motion management strategy by adopting AlignRT system. The purpose of this study is to develop, validate, and prove the feasibility of VRT-DIBH technique for lung and liver SBRT.

Conditions

Interventions

TypeNameDescription
DEVICETidal volume measured by spirometer and DIBH surface from CT imageThe SOC DIBH CT scan will be transferred to Eclipse treatment planning system for treatment planning. Tidal volume measured by spirometer and DIBH surface from CT image will be used as a reference surface for patient treatment initial setup. CBCT will be acquired before each fractional treatment for the patient alignment before beam delivery.

Timeline

Start date
2019-03-01
Primary completion
2021-07-20
Completion
2023-06-21
First posted
2020-12-28
Last updated
2024-11-25
Results posted
2024-11-25

Locations

1 site across 1 country: United States

Source: ClinicalTrials.gov record NCT04686500. Inclusion in this directory is not an endorsement.