Clinical Trials Directory

Trials / Completed

CompletedNCT04236973

Truenat™ HCV RNA Assay Evaluation

Multicentre Clinical Performance Evaluation of the Molbio Diagnostics Truenat™ HCV RNA Assay

Status
Completed
Phase
N/A
Study type
Interventional
Enrollment
1,330 (actual)
Sponsor
Foundation for Innovative New Diagnostics, Switzerland · Academic / Other
Sex
All
Age
16 Years
Healthy volunteers
Accepted

Summary

Molbio Diagnostics Ltd. (India) has developed the Truelab™ Real Time quantitative PCR system that is widely used in India for diagnostics of tuberculosis (TB). The system consists of two portable machines and two microfluidic cartridges and can be used in point-of-care settings. The manufacturer has recently developed a new assay that detects HCV RNA, the Truenat™ HCV RNA assay. It is a simple two-step assay for RNA extraction and amplification with a total turnaround time of 60 min, using whole blood, plasma and serum as sample types. Most importantly, the assay can be performed from a drop of capillary blood eliminating the need for venous access and blood draw and increasing its usability in the settings where phlebotomy service are not available. To date, validation of the assay was performed using archived plasma specimens and contrived whole blood specimens. FIND aims to conduct a multicentre evaluation to assess the assay's sensitivity, specificity and quantitative accuracy in freshly collected whole blood, plasma and serum specimens from target populations. The evaluation aims to gather performance data in line with the requirements set forth in the Common Technical Specifications 2009/886/EC (CTS) of the CE In Vitro Diagnostics Medical Devices Directive 98/79/CE (CE-IVDD), as well as the World Health Organization (WHO) Technical Specification Series 10 (draft) (TSS-10) for In vitro diagnostic (IVDs) medical devices used for the qualitative and quantitative detection of HCV RNA.

Detailed description

Even though hepatitis C virus (HCV) infection can now be cured, globally, fewer than 20% of the 71 million living with HCV are aware of their status; diagnosis remains a major challenge. While serology screening tests that detect the presence of antibodies against HCV, can be performed in decentralized settings, these tests cannot distinguish between resolved (about 25% of seropositive cases) and active infection. Therefore, positive serology test results require further confirmation by laboratory-based methods that detects the viral particle. The two-step testing process results in high drop-off in confirmatory testing among HCV seropositive individuals, especially in low- and middle-income countries (LMIC) where laboratory infrastructure is lacking. Providing access to confirmatory testing in decentralized settings has a high potential to increase the coverage of HCV testing services in LMIC. Molbio Diagnostics Ltd. (India) has developed the Truelab™ Real Time quantitative PCR system that is widely used in India for diagnostics of tuberculosis (TB). The system consists of two portable machines and two microfluidic cartridges and can be used in point-of-care settings. The manufacturer has recently developed a new assay that detects HCV RNA, the Truenat™ HCV RNA assay. It is a simple two-step assay for RNA extraction and amplification with a total turnaround time of 60 min, using whole blood, plasma and serum as sample types. Most importantly, the assay can be performed from a drop of capillary blood eliminating the need for venous access and blood draw and increasing its usability in the settings where phlebotomy service are not available. To date, validation of the assay was performed using archived plasma specimens and contrived whole blood specimens. FIND aims to conduct a multicentre evaluation to assess the assay's sensitivity, specificity and quantitative accuracy in freshly collected whole blood, plasma and serum specimens from target populations. The evaluation aims to gather performance data in line with the requirements set forth in the Common Technical Specifications 2009/886/EC (CTS) of the CE In Vitro Diagnostics Medical Devices Directive 98/79/CE (CE-IVDD), as well as the World Health Organization (WHO) Technical Specification Series 10 (draft) (TSS-10) for In vitro diagnostic (IVDs) medical devices used for the qualitative and quantitative detection of HCV RNA. The data will be published as a peer-reviewed report and made available to the test manufacturer to support regulatory submissions.

Conditions

Interventions

TypeNameDescription
DEVICEMolbio Truenat™ HCV assay (the investigational product)The Molbio Truenat™ HCV assay (the investigational product) is a quantitative chip-based Real Time Reverse Transcription Polymerase Chain Reaction (qRT-PCR) tests for the detection of HCV genomic RNA from human whole blood, EDTA plasma and serum. The Truenat™ HCV assay runs on the Truelab™ Uno, Truelab™ Duo or Truelab™ Quadro Dx. Results of testing by Truenat™ HCV assay will not be used to make any clinical decisions, and should not be communicated to study participants.
OTHERAbbott RealTime HCV assayPlasma specimens will be also tested on Abbott RealTime HCV assay that is approved for HCV diagnostics use by countries' authorities. Only the results of Abbott RealTime HCV assay may be used to make clinical decisions.

Timeline

Start date
2020-01-12
Primary completion
2021-07-20
Completion
2021-07-20
First posted
2020-01-22
Last updated
2021-07-26

Locations

5 sites across 5 countries: Denmark, Ethiopia, Georgia, Thailand, Ukraine

Source: ClinicalTrials.gov record NCT04236973. Inclusion in this directory is not an endorsement.