Clinical Trials Directory

Trials / Completed

CompletedNCT04209972

The Benefit of UHR-CT: Precision of Repeated Volume Measurements of Pulmonary Nodules

The Benefit of Ultra-high Resolution Computed Tomography: Precision of Repeated Volume Measurements of Pulmonary Nodules

Status
Completed
Phase
N/A
Study type
Interventional
Enrollment
92 (actual)
Sponsor
Radboud University Medical Center · Academic / Other
Sex
All
Age
18 Years
Healthy volunteers
Not accepted

Summary

To assess the variability of semi-automated volume measurements of pulmonary nodules on same-day repeated scans of equal radiation dose from two different CT scanners: One high-end CT scanner with standard spatial resolution (CT1) and one UHRCT scanner (CT2), in patients with known or suspected pulmonary metastases.

Detailed description

Rationale: Ultra-high resolution computed tomography (UHRCT) produces radiological images with a spatial resolution of 0.25 mm in a matrix of 1024x1024. This should decrease measurement variation of nodule growth as a marker of malignancy, by making nodule delineation more precise for automatic volumetry segmentation and volume doubling time assessment than in conventional CT. If possible, this can shorten follow up of incidental pulmonary nodules to exclude malignancy, with less medicalisation and patient anxiety. Objective: To assess the variability of semi-automated volume measurements of pulmonary nodules in patients with known or suspected pulmonary metastases on same-day repeated scans of equal radiation dose from two different CT scanners: CT scanner with standard spatial resolution (conventional CT, CT1) and UHRCT (CT2). Study design: This is a single center prospective trial on 80 patients with known or suspected pulmonary metastases who are scheduled for chest and/or abdominal CT. Study participants will undergo two additional partial chest CT scans on either CT 1 or CT 2 for research purpose only, at similar radiation dose. Patients are equally divided across CT 1 and 2. Study population: Patients who are 18 years or older with known solid pulmonary nodules compatible with metastases and who are willing and able to give informed consent are eligible. Patients are excluded if they have less than two eligible pulmonary nodules with a z range of 16 cm. Nodules with calcifications, surrounding opacities, or vessel- or pleural abutment will be excluded from analysis. Main study parameters/endpoints: The main endpoint of this study is the upper limit of the 95% confidence interval of repeated semi-automated nodule volume measurements of both CT scanners. Nature and extent of the burden and risks associated with participation, benefit and group relatedness: CT imaging is associated with risks related to the use of ionizing radiation. The CT protocol including the scheduled CT scan and the additional scans in this study has been carefully designed to have a total radiation dose at the same level as the achievable diagnostic reference level of chest CT in the Netherlands, which is 542 mGycm in 2013 (1). The burden associated with the two extra study CT acquisitions comprises a dose length product (DLP) of 120.4 mGy•cm for research (1.7 mSv, which is lower than the background radiation of one year in the Netherlands, with a conversion factor of 0.014 from Deak et al (2)).

Conditions

Interventions

TypeNameDescription
DEVICEPrecision Ultra-High-Resolution CT scannerPatients will be divided on one of the two CT scanners. This group will be scanned on the Precision CT scanner.
DEVICEGenesis high-end CT scannerPatients will be divided on one of the two CT scanners. This group will be scanned on the Aquilion one Genesis.

Timeline

Start date
2019-03-11
Primary completion
2019-07-30
Completion
2019-07-30
First posted
2019-12-24
Last updated
2019-12-24

Locations

1 site across 1 country: Netherlands