Trials / Completed
CompletedNCT04148664
A Trial of High Power-Short Duration Versus Standard Power-Long Duration Radiofrequency Ablation for Treatment of Atrial Fibrillation
SHORT-AF: A Randomized Trial of High Power-Short Duration Versus Standard Power-Long Duration Radiofrequency Ablation for Treatment of Atrial Fibrillation
- Status
- Completed
- Phase
- N/A
- Study type
- Interventional
- Enrollment
- 60 (actual)
- Sponsor
- University of California, San Francisco · Academic / Other
- Sex
- All
- Age
- 18 Years
- Healthy volunteers
- Not accepted
Summary
This study is a prospective, randomized controlled study to compare overall clinical outcomes between High Power Short Duration (HPSD) and standard radiofrequency (RF) ablation settings for Atrial Fibrillation (AF) ablation in the treatment of subjects with paroxysmal or persistent Atrial Fibrillation.
Detailed description
Pulmonary vein isolation using radiofrequency (RF) ablation is a widely used treatment strategy for atrial fibrillation. Peri-procedural complications rates are estimated at between 1.5 - 6% with the two most feared complications being stroke and atrial-esophageal fistula. The risk of these complications increases with (1) longer left atrial dwell times with greater potential for clot formation and (2) esophageal heating during delivery of radiofrequency energy. 'High-power short-duration' (HPSD) is an increasingly utilized strategy to decrease procedure duration and minimize the risk of these complications. Potential mechanisms for benefit include: (1) shorter left atrial dwell times due to more efficient lesion delivery (2) rapid, but more controlled, resistive tissue heating, which avoids deeper, passive conductive heating that can reach the esophagus and cause injury. Moreover, HPSD may improve lesion efficacy and durability due to less recover of excitability after ablation. Despite being widely used and several retrospective studies, there have been no prospective randomized trials comparing outcomes between HPSD and standard RF ablation settings. Moreover, the investigators are interested to see if a shorter procedure (less time in the left atrium) is associated with a lower rate of silent cerebral infarctions (which are occasionally seen on MRI brain post RF ablation - their clinical significance is unclear).
Conditions
Interventions
| Type | Name | Description |
|---|---|---|
| PROCEDURE | Catheter ablation with Standard RF ablation settings | Standard RF ablation settings (n=30): Maximum 35W for 20-40 seconds, per operator standard. |
| PROCEDURE | Catheter ablation with High Power Short Duration RF ablation settings | High power short duration RF (n=30): Maximum 50 Watts for 5-15 seconds, per operator standard. |
Timeline
- Start date
- 2019-12-03
- Primary completion
- 2020-12-01
- Completion
- 2022-12-30
- First posted
- 2019-11-01
- Last updated
- 2023-08-16
Locations
1 site across 1 country: United States
Source: ClinicalTrials.gov record NCT04148664. Inclusion in this directory is not an endorsement.