Subcostal Temporary Extracardiac Pacing Study

Subcostal Temporary Extracardiac Pacing Study: Clinical Study of the AtaCor Extracardiac Pacing System

Status
Completed
Phase
N/A
Study type
Interventional
Enrollment
12 (actual)
Sponsor
AtaCor Medical, Inc. · Industry
Sex
Male
Age
18 Years
Healthy volunteers
Not accepted

Summary

Up to 15 subjects will be enrolled to evaluate initial safety and performance of the AtaCor Temporary Pacing System. Safety will be evaluated through analysis of all Adverse Events. Performance will be evaluated through (1) the incidence of successful StealthTrac Lead placement using the MACH I Delivery Tool, (2) electrical performance measurements, such as pacing capture thresholds, sensed R-wave amplitudes, pacing impedance and skeletal muscle stimulation at rest. Appropriate sensing and pacing capture will be ascertained from ECG Holter Monitor recordings during periods of in-hospital ambulation.

Detailed description

Enrolled Subjects will have both a market-released transvenous pacing lead and investigational StealthTrac Temporary Pacing Lead placed and evaluated. During the placement procedure, diagnostic echocardiograms will be performed to detect any new pericardial effusions. Once inserted, the StealthTrac Lead will be used to deliver high rate pacing (160 to 220 BPM) in clinically appropriate Subjects, to demonstrate suitability for use in transcatheter procedures that require brief periods of temporary pacing. Lead positions will be documented using fluoroscopic images before the Subject leaves the procedure room. On the following two days after the procedure, electrical performance will be evaluated in multiple postures and appropriate pacing and sensing will be confirmed during activity using ECG Holter monitor recordings. Subjects will also be asked patient-centric questions about their experience with the StealthTrac Lead throughout the study period. Evaluations will continue for a minimum of two (2) and a maximum of seven (7) days after the index procedure. Prior to StealthTrac Lead removal, an X-Ray will be taken to document the final lead position and diagnostic echocardiography will be performed to (1) detect any latent pericardial effusions and (2) assess differences in cardiac function with intrinsic conduction and while pacing. A final follow-up will be performed 25-30 days after removal to identify any latent adverse events before the Subject exits the study.

Conditions

Interventions

TypeNameDescription
DEVICESubcostal Temporary Extracardiac Pacing LeadThe StealthTrac Lead is designed to facilitate extracardiac temporary ventricular pacing and sensing. The distal end of the StealthTrac Lead is designed to reside within the connective tissue of the anterior mediastinum outside the pericardium. The StealthTrac Lead is delivered using the MACH I Delivery Tool, which is designed to facilitate insertion of the StealthTrac Lead through a small skin incision parallel to the sternum, without the need for fluoroscopic guidance.

Timeline

Start date
2019-08-25
Primary completion
2019-11-07
Completion
2019-11-07
First posted
2019-09-19
Last updated
2021-02-03
Results posted
2021-02-03

Locations

1 site across 1 country: Paraguay

Regulatory

Source: ClinicalTrials.gov record NCT04096365. Inclusion in this directory is not an endorsement.