Capillary Blood Volume and Pain Perception Obtained in a Process of Puncturing With Different Safety Lancets

A Randomized, Single Blind, Uni-centre Study in Healthy Volunteers, to Determine Capillary Blood Volume and Pain Perception Obtained in a Process of Puncturing With Different Safety Lancets

Status
Completed
Phase
N/A
Study type
Interventional
Enrollment
300 (actual)
Sponsor
HTL-Strefa S.A. · Industry
Sex
All
Age
18 Years – 65 Years
Healthy volunteers
Accepted

Summary

The study was a randomized, single blind, uni-center pivotal study in male and female healthy volunteers, to determine capillary blood volume and pain perception obtained in a process of puncturing with different safety lancets. A group of three hundred (300) Caucasian male and female Subjects was randomized (aged \> 18 to \< 65 years old) who were in good physical and mental condition. The Subjects had 4 (four) fingers (two fingers from each hand) lanced each with a different type of safety lancet. Overall, 4 (four) different safety lancets were investigated per each Subject. Overall, 48 (forty eight) different safety lancets (4 different safety lancets per one Subject) were investigated, the capillary blood volume and pain were measured.

Conditions

Interventions

TypeNameDescription
DEVICEActi-Lance LiteFinger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICEActi-Lance UniversalFinger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICEActi-Lance SpecialFinger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICEMedlance Plus Super LiteFinger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICEMedlance Plus LiteFinger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICEMedlance Plus UniversalFinger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICEMedlance Plus ExtraFinger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICEMedlance Plus SpecialFinger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICEProlance Micro FlowFinger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICEProlance Low FlowFinger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICEProlance Normal FlowFinger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICEProlance High FlowFinger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICEProlance Max FlowFinger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICEProlance PediatricFinger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICEHaemolance Micro FlowFinger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICEHaemolance Low FlowFinger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICEHaemolnace Normal FlowFinger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICEHaemolance High FlowFinger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICEHaemolance Max FlowFinger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICEHaemolance PediatricFinger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICEMediSafe Solo 29 GaugeFinger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICEMediSafe Solo 23 GaugeFinger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICEergoLance Micro FlowFinger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICEergoLance Normal FlowFinger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICEergoLance High FlowFinger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICEMedlance (light blue)Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICEMedlance (blue)Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICEMedlance (dark blue)Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICEMedlance (yellow)Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICEMedlance (orange)Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICEMedlance (red)Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICEUnistik 3 ExtraFinger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICEUnistik 3 ComfortFinger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICEUnistik TouchFinger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICEReadyLance (blue)Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICEReadyLance (orange)Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICEAssureLanceFinger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICEAssureLance MicroFlowFinger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICESurgilanceFinger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICESurgilance Lite (grey)Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICESurgilance Lite (purple)Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICESteriLance Lite2Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICESensiLance (pink)Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICESteriLance PressFinger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICESensiLance (orange)Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICEmylifeFinger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICEVeriFineFinger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
DEVICEMicrodotFinger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol. Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Timeline

Start date
2017-12-08
Primary completion
2017-12-17
Completion
2018-01-30
First posted
2018-11-23
Last updated
2018-11-26

Locations

1 site across 1 country: Poland

Source: ClinicalTrials.gov record NCT03752229. Inclusion in this directory is not an endorsement.