Trials / Completed
CompletedNCT03614416
Clinical Trial Testing the BioSerenity Pulse Oximeter (EOXY) Efficacy, Performance and Safety in Healthy Volunteer
- Status
- Completed
- Phase
- N/A
- Study type
- Interventional
- Enrollment
- 13 (actual)
- Sponsor
- BioSerenity · Industry
- Sex
- All
- Age
- 18 Years
- Healthy volunteers
- Accepted
Summary
The purpose of the study is to test the performance and safety of the Bioserenity pulse oxymeter EOXY. For that, and in agreement with ISO60601-2-61 norm and FDA regulation, measures of heart rate, with a gold-standard, and SaO2 measures generated by blood samples will be compared with Sp02 and heart rate measures that are generated by our EOXY. Thirteen healthy subjects will participate in the study in total. Three subjects will be enrolled during a primary phase .They will participate at a secondary phase as well as ten others subjects. The purpose of the primary phase is to test the body area where the measurement can be done. The purpose of the secondary phase is to measure SpO2 and HR in desaturation conditions.
Detailed description
The primary objective is to evaluate and document the SpO2 accuracy of the pulse oximeter equipment, evaluating the performance of the device compared to invasive method blood sampling (SaO2) by CO-Oximeter in healthy subjects during invasive controlled desaturation study at different ranges (100 % to 70% SaO2). Subject safety during desaturation and saturation period, will be evaluated by reviewing vital sign: heart rate (HR) continuously and blood pressure/temperature between each desaturation plateau. There are two secondaries objectives, the measurement of the variation of the fraction of inspired oxygen (FiO2), and the validation of the algorithm for data review and analysis. For the study primary phase, the purpose of this phase is to test the body area where the SpO2 is better detected and define 3 measurement areas. For this, three healthy volunteers will be placed in normal O2 saturation, the SpO2 will be measured using the Bioserenity oximeter (noninvasive method) and will be compared to a gold-standard oximeter CE-marked. The investigator will determine the best area in the body where the SpO2 could be taken for the following subjects. During a second phase, the measurement of Sp02 and HR with EOXY, will be performed in O2 saturation conditions up to 100% and then in desaturation by plateau from 100 % to 70%. 3 EOXY pulse oximeter recorded the SpO2 at the same time in 3 different areas of the body. As required by the standard, Sp02 measures will be simultaneously compared to SaO2 values measured by blood sampling taken from an indwelling arterial catheter. The HR obtained with EOXY will be compared simultaneously to a gold standard oximeter CE-marked. As the example in the standard, it was choosen to do the test on the 70-100% range, with 5 plateaus (100%-97%, 97%-92%, 92%-85%, 84%-78%, 77%-70%) and to collect five Sp02 and Sa02 values (blood samples) and five HR measures, per plateau, for each of 13 subjects. The desaturation will be controlled thanks' the FiO2 measures, that should allow to bring subjects near target levels of desaturation plateau.
Conditions
Interventions
| Type | Name | Description |
|---|---|---|
| DEVICE | EOXY device | Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard |
| DEVICE | Gold standard oximeter | Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard |
| PROCEDURE | SaO2 sampling | Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard |
Timeline
- Start date
- 2017-12-08
- Primary completion
- 2017-12-15
- Completion
- 2018-11-06
- First posted
- 2018-08-03
- Last updated
- 2021-01-12
- Results posted
- 2021-01-12
Locations
1 site across 1 country: Belgium
Source: ClinicalTrials.gov record NCT03614416. Inclusion in this directory is not an endorsement.