Clinical Trials Directory

Trials / Completed

CompletedNCT03200457

Comparison of the Variability of Hepatic Load Quantification in Iron and Fat Estimated by MRI at 1.5 and 3 Tesla

Status
Completed
Phase
N/A
Study type
Interventional
Enrollment
92 (actual)
Sponsor
Rennes University Hospital · Academic / Other
Sex
All
Age
18 Years
Healthy volunteers
Not accepted

Summary

The study proposes to compare the results of two examinations at 1.5 and 3 Tesla obtained on the same day on a series of 80 patients.

Detailed description

Quantification of hepatic iron by MRI is validated at 1.5 Tesla on a small number of subjects with some discrepancies between the two most used techniques (liver to muscle ratio or calculation of the decay rate of the T2 \* signal). Since the initial publications at 1.5 Tesla, new technical innovations have emerged which can allow better quantification. Recently, the investigators have applied these innovations to Rennes and have validated (SURFER study) the possibility of making a quantification of iron and fat on two 3 Tesla devices with good results in a single apnea acquisition. But it is not uncommon to have, in a structure, different magnetic field devices, and it is important to ensure that the results are transposable. It is also crucial for multicentre or cohort studies to ensure consistency. It is therefore proposed to compare the results of two examinations at 1.5 and 3 Tesla obtained on the same day on a series of 80 patients.

Conditions

Interventions

TypeNameDescription
DEVICE1,5 Tesla abdominal MRIMeasurement of the fat and iron
DEVICE3 Tesla abdominal MRIMeasurement of the fat and iron

Timeline

Start date
2018-07-20
Primary completion
2020-08-28
Completion
2020-08-28
First posted
2017-06-27
Last updated
2022-02-15

Locations

2 sites across 1 country: France

Source: ClinicalTrials.gov record NCT03200457. Inclusion in this directory is not an endorsement.