Clinical Trials Directory

Trials / Unknown

UnknownNCT03074266

Clinical Performance of the Investigational GEM Hemochron® 100 in Populations Ages 18 Years and Older

Status
Unknown
Phase
Study type
Observational
Enrollment
140 (estimated)
Sponsor
Accriva Diagnostics · Industry
Sex
All
Age
18 Years
Healthy volunteers
Accepted

Summary

This study is designed to evaluate and verify the clinical accuracy of the Hemochron® MCS 7000 Whole Blood Microcoagulation System; a point-of-care, whole blood Activated Clotting Time (ACT) measurement in comparison with the Hemochron Signature Elite® Whole Blood Microcoagulation System; a 510(k) cleared predicate device. This is a multi-center, prospective method comparison, with data to be analyzed separately from each site as well as combined for all sites.

Detailed description

The study will be performed in the clinical setting by trained professional operators. Group 1: Whole Blood Coagulation Test 1, GEM Hemochron® 100 Group 2: Whole Blood Coagulation Test 2, Hemochron® Signature Elite® 7000 Whole Blood Microcoagulation System Ex-vivo whole blood samples destined for discard after all medically directed tests are complete will be used from heparinized patients undergoing standard of care interventional procedures in different clinical settings. These include, but are not limited to, the cardiovascular operating room (CVOR), cardiac catheterization lab, surgical intensive care unit (SICU), the electrophysiology (EP) lab, and in extracorporeal membrane oxygenation (ECMO) applications. A distribution of ACT values is necessary to evaluate clinical accuracy. The target number of subjects for each site is 100. Samples will be examined at baseline (pre-heparin administration) and at the end of the procedure. Heparinized samples may be obtained for ACT testing once or multiple times while heparin is administered throughout the duration of the procedure. The frequency and time intervals of ACT testing will be determined by the Standard of Care. No additional blood volume or extra blood draws will occur as the result of this study. The work flow for the testing process will be consistent across all sites. Expected duration of the study is dependent upon the site logistics but generally is expected to last no more than four to six weeks. ACT results generated from the GEM Hemochron 100 instrument during this study will be used solely for research purposes and will not be used for anticoagulation management of study subjects. No patient follow up is required. The study will include two GEM Hemochron 100 instruments, trained operators, and two Signature Elite instruments for each procedure. Guidance was obtained from Point-of-Care Monitoring of Anticoagulation Therapy; Approved Guideline (CLSI POCT14-A), CLSI EP09-A3: "Method Procedure Comparison and Bias Estimation Using Patient Samples; Approved guideline Third Edition" and peer reviewed literature. All study data will be collected on Case Report Forms and all analysis will be performed by Accriva.

Conditions

Interventions

TypeNameDescription
DIAGNOSTIC_TESTGEM Hemochron 100 ACTActivated Clotting Time

Timeline

Start date
2017-06-08
Primary completion
2018-12-01
Completion
2018-12-01
First posted
2017-03-08
Last updated
2018-07-26

Locations

9 sites across 1 country: United States

Regulatory

Source: ClinicalTrials.gov record NCT03074266. Inclusion in this directory is not an endorsement.