Summary

The goal of this study is to compare the effectiveness of pacing from a physiologic His bundle (HB) lead position versus with the standard coronary sinus (CS) lead position in subjects with heart failure undergoing cardiac resynchronization therapy (CRT). While placement of left ventricular leads via the coronary sinus has anatomic limitations, we hypothesis that the achievement of QRS narrowing with His bundle capture will be superior for improving systolic function by echocardiographic indices (ejection fraction and strain) and quality of life and decreased rehospitalization and mortality.

Detailed description

This is a randomized, single-blinded study of 40 patients to a strategy of HB pacing versus CS pacing, and remain blinded to their treatment allocation. Both treatment options use standard-of-care, FDA-approved devices. The distinction is only in the allocation toward HB pacing and CS pacing. Treating physicians will be aware of assignment in order to facilitate routine device follow-up. Echocardiographic and electrocardiographic evaluation will also be performed in a blinded manner. Cross-over is permitted between treatment group allocation if: * CS lead cannot be placed due to difficult cannulation of the CS, limited branches at the posterolateral or lateral wall, or phrenic nerve capture. These subjects may then cross-over to HB pacing. * HB pacing subjects may cross-over if HB lead cannot be positioned with adequate stability and reasonable pacing output, or if QRS width does not narrow by at least 20% or to a QRS width of ≤ 130 msec. Implant procedure will be per routine percutaneous access, as is standard for pacemaker and ICDs. All subjects will receive an FDA-approved cardiac resynchronization therapy pacemaker or defibrillator device, as per standard of care outlined for the subject. In order to facilitate optimal lead placement, arterial access for levo-phase CS angiography and/or LV septal mapping to characterize site of bundle-branch block may also be performed, at the discretion of the implanting physician per his/her standard practice. Follow-up will be performed at 2 weeks post-implant for incision check and device interrogation as is standard of care. In addition, routine device and clinical follow-up will be scheduled at 1, 3, 6, and 12 months. Electrocardiography (ECG) will be performed pre-implant, prior to hospital discharge, at 3 months, 6 months, and 12 months. Echocardiography will be performed pre-implant and 6 months to evaluate for change in LVEF, chamber dimension, and wall motion with strain imaging as is standard of care in the treatment of patients with advanced heart failure. NYHA functional class and quality of life (utilizing the Kansas City Cardiomyopathy Questionnaire) will be assessed pre-implant and at 6 months.

Conditions

Interventions

TypeNameDescription
DEVICECRT PacemakerCardiac Resynchronization Therapy (CRT) is the use of a pacemaker with two endocardial leads placed in the right atrium (RA) and right ventricle (RV). The third lead is traditionally placed in a tributary of the coronary sinus (CS) overlying the epicardial surface of the left ventricle (LV). Alternatively, the third lead may be positioned based on mapping of the common His bundle and actively fixed to achieve QRS normalization via direct His bundle capture.

Timeline

Start date
2016-05-17
Primary completion
2020-07-31
Completion
2020-07-31
First posted
2016-03-07
Last updated
2021-09-05
Results posted
2021-08-10

Locations

8 sites across 1 country: United States

Regulatory

Source: ClinicalTrials.gov record NCT02700425. Inclusion in this directory is not an endorsement.