Clinical Trials Directory

Trials / Completed

CompletedNCT02511652

Ambu AuraGain vs LMA Supreme in Laparoscopic Surgery

Comparison of the Ambu AuraGain vs. LMA Supreme in Paralysed Patients Undergoing Gynecologic Laparoscopic Surgery in the Trendelemburg Position

Status
Completed
Phase
N/A
Study type
Interventional
Enrollment
60 (actual)
Sponsor
Hospital Clinic of Barcelona · Academic / Other
Sex
Female
Age
18 Years – 80 Years
Healthy volunteers
Not accepted

Summary

The aim of the study is to compare the performance of two supraglottic airway devices (SGA), the new Ambu AuraGain and the LMA Supreme, in sixty female patients undergoing gynaecologic laparoscopy in the trendelemburg position. Primary outcome is the airway seal pressure. Secondary outcomes are ease of insertion, quality of ventilation, endoscopic view of the glottis, gastric tube passage and complications.

Detailed description

The patients are randomly assigned to 2 groups: * Group 1: Ambu AuraGain. * Group 2: LMA Supreme, Teleflex Procedure: The investigation protocol contains the following sections: 1. Induction of anaesthesia. Intravenous infusion of remifentanil and propofol targeting the effect site (2-4 ng/ml and 4-6 µg/ml respectively). No muscle relaxant will be used for insertion of the SGA. Rocuronium 0.2 mg/Kg will be administered before initiation of pneumoperitoneum. 2. Insertion of the SGAs. The size of the SGA device used is based on the manufacturers' recommendations. All devices are deflated a lubricated prior to use. Once inserted, the cuff is be inflated with a manometer up to 60 cm H20 Position of the device is adjusted if needed. Data recorded: size of SGA, time an number of attempts. 3. Fibreoptic evaluation of the SGAs anatomical position: complete view of the vocal cords (I), epiglottis visible inside, but not causing obstruction (II), epiglottis visible and obstructing the glottic inlet (III), or glottic structures not identified (IV). 4. Functionality of the gastric drainage channel of the SGAs: passage of a 16 G size tube. 5. Measurement of airway seal pressure (oropharyngeal leak pressure (OLP): at baseline, and at 15, 30 and 60 minutes. The maximum pressure allowed is 40 cm H2O. 6. Ventilatory mechanics and parameters are measured at baseline, and at 15, 30 and 60 minutes. Perioperative complications: Hiccup, gastric distension, regurgitation / Aspiration, airway obstruction, laryngospasm, dental, mucosal or tongue injury, hypoxia (SpO2 \< 92%) 7. Removal of the SGAs: Presence of blood - 3 level grading (+/++/+++) Postoperative complications: sore throat, hoarseness, dysphonia, dysphagia, dysphagia: 3-point scale

Conditions

Interventions

TypeNameDescription
DEVICEAmbu AuraGain evaluationEvaluation of clinical performance in terms of Insertion, ventilation, position and complications
DEVICELMA Supreme evaluationEvaluation of clinical performance in terms of Insertion, ventilation, position and complications
PROCEDURELaparoscopic surgeryGynecologic laparoscopic surgery in the trendelemburg position

Timeline

Start date
2015-06-01
Primary completion
2015-12-01
Completion
2015-12-01
First posted
2015-07-30
Last updated
2016-02-22

Locations

1 site across 1 country: Spain

Source: ClinicalTrials.gov record NCT02511652. Inclusion in this directory is not an endorsement.