Clinical Trials Directory

Trials / Completed

CompletedNCT02352298

Protocol for the Evaluation of Dario™ BGMS Performance at Elevated Altitude

Protocol for the Evaluation of Dario™ Blood Glucose Monitor (BGMS) Performance at Elevated Altitude

Status
Completed
Phase
N/A
Study type
Interventional
Enrollment
52 (actual)
Sponsor
LabStyle Innovations Ltd. · Industry
Sex
All
Age
18 Years
Healthy volunteers
Accepted

Summary

The purpose of this study is to assess the performance of the Dario Blood Glucose Monitor (BGMS), at elevated altitudes, by obtaining whole blood fingerstick samples from individuals and analyzing the blood on the Dario Blood Glucose Monitor (BGMS) and comparing the readings to readings obtained from the YSI 2300 STAT Plus glucose analyzer.

Detailed description

The study will be performed with 50 fresh capillary blood samples, collected by fingerstick using Dario single use lancet. Additional fingerstick with dedicated disposable lancets will be used to collect sufficient capillary fresh blood to be analyzed by the YSI 2300 STAT Plus analyzer. Additional fingersticks may be necessary if volume of blood from first fingerstick is insufficient.

Conditions

Interventions

TypeNameDescription
DEVICEDario Blood Glucose Monitoring SystemFingerstick to obtain blood sample for analysis with the Dario BGMS
DEVICEYSI AnalyzerFingerstick to obtain blood sample for analysis with YSI

Timeline

Start date
2015-02-01
Primary completion
2015-02-01
Completion
2015-02-01
First posted
2015-02-02
Last updated
2017-06-09
Results posted
2017-06-09

Locations

1 site across 1 country: United States

Source: ClinicalTrials.gov record NCT02352298. Inclusion in this directory is not an endorsement.