Clinical Trials Directory

Trials / Completed

CompletedNCT02266407

Surgical / Economic Effect of the Aquamantys System in Blood Management for Aseptic and Septic Revision TKA

The Surgical and Economic Effect of the Aquamantys System in Blood Management During and Following Aseptic and Septic Revision TKA

Status
Completed
Phase
N/A
Study type
Interventional
Enrollment
80 (actual)
Sponsor
Medtronic Surgical Technologies · Industry
Sex
All
Age
18 Years
Healthy volunteers
Accepted

Summary

To assess the surgical and economic effectiveness of the Aquamantys System in managing intra- and post-operative blood loss, and reducing transfusion needs in patients undergoing both aseptic and septic revision total knee arthroplasty.

Detailed description

The use of the saline-coupled bipolar sealing technology (Aquamantys® System, Medtronic Advanced Energy, LLC, Portsmouth, NH, USA) represents a modern approach to reducing perioperative blood loss in patients undergoing total joint arthroplasty. Unlike standard electrocautery, which use monopolar radiofrequency energy, this technology uses bipolar radiofrequency energy combined with a continuous-flow saline at the electrode tip to prevent tissue temperatures from exceeding 100 °C thus minimizing tissue charring. This eliminates the smoke and eschar formation seen with standard electrocautery. The temperature achieved is sufficient to induce a process called Transcollation(TM) which shrinks the collagen fibers in the walls of blood vessels, effectively sealing the blood vessels up to 1mm in diameter, resulting in the reduction of bleeding from soft tissue. Suction is used to remove the saline from the surgical field with the treated tissue turning a light tan color. With normal use, the depth of penetration of this effect is typically less than 2mm. The technology is cleared by the US FDA (510(k)) and EU (CE Mark) for use on soft tissue \& bone and will be used according to its intended packaging indications for use. The current assessment proposes a technique in which a disposable Aquamantys® hand piece is used to treat targeted areas which are bleeding or expected to bleed during surgery. This postmarket study will assess the surgical and economic effectiveness of Aquamantys® System in managing intra-operative blood loss and transfusion rates in patients undergoing septic and aseptic revision TKA.

Conditions

Interventions

TypeNameDescription
PROCEDUREAquamantys SystemThe use of the saline-coupled bipolar sealing technology (Aquamantys® System, Medtronic Advanced Energy, LLC, Portsmouth, NH, USA) uses bipolar radiofrequency energy combined with a continuous-flow saline at the electrode tip to prevent tissue temperatures from exceeding 100 °C thus minimizing tissue charring. The temperature achieved is sufficient to induce a process called Transcollation(TM) which shrinks the collagen fibers in the walls of blood vessels, effectively sealing the blood vessels up to 1mm in diameter, resulting in the reduction of bleeding from soft tissue.

Timeline

Start date
2012-06-01
Primary completion
2012-09-01
Completion
2012-09-01
First posted
2014-10-17
Last updated
2020-02-27

Locations

1 site across 1 country: Germany

Source: ClinicalTrials.gov record NCT02266407. Inclusion in this directory is not an endorsement.