Clinical Trials Directory

Trials / Completed

CompletedNCT02094209

Focused Registry SmartFix

Clinical Data Collection With a Novel Biofeedback Technology for Continuous Monitoring of Bone Healing

Status
Completed
Phase
Study type
Observational
Enrollment
10 (actual)
Sponsor
AO Clinical Investigation and Publishing Documentation · Academic / Other
Sex
All
Age
18 Years
Healthy volunteers
Not accepted

Summary

20 patients having received an AO large external fixator at the tibia will be equipped with a data logger device (AO Fracture Monitor) attached post-operation to a connection rod of the external fixator. The device continuously measures deformation of the fixator frame due to weight bearing for up to 6 months by means of a strain gauge. Several parameters are calculated from the recorded change in the strain signal and are stored at a regular interval.

Detailed description

Maturation of fracture callus leads to unloading of the fracture fixation hardware due to an increased load-share of the repair tissue. A novel data logger device (AO Fracture Monitor) continuously measures the decline in fixation hardware deflection under physiological loading as indirect indicator for the healing progress. Parameters obtained from the data logger device carry potential to significantly improve the assessment of fracture healing in the future. Meaningful interpretation of measurements requires a set of clinical reference data. 20 patients having received an AO large external fixator at the tibia will be equipped with a data logger device (AO Fracture Monitor) attached post-operation to a connection rod of the external fixator. The device continuously measures deformation of the fixator frame due to weight bearing for up to 4 months by means of a strain gauge. Several parameters are calculated from the recorded change in the strain signal and are stored at a regular interval. Data is collected from the AO Fracture Monitor at follow-up visits of the patient by wireless data transfer. Together with additional variables such as treatment details, fracture healing and pain reported by the patient, the collected data is used to build up a database. Data from the AO Fracture Monitor will be correlated with patient data to investigate the relevance and reliability of the data derived from the AO Fracture Monitor. In this phase, the study does not imply changes on the operational treatment, nor does it allow for therapeutic consequences based on the derived data.

Conditions

Interventions

TypeNameDescription
DEVICESmartFixAO Fracture Monitor attached postoperatively to a connection rod of the external fixator. The device continuously measures deformation of the fixator frame due to weight bearing for up to 4 months by means of a strain gauge.

Timeline

Start date
2015-01-01
Primary completion
2017-02-01
Completion
2017-07-01
First posted
2014-03-21
Last updated
2020-08-12

Locations

1 site across 1 country: Germany

Source: ClinicalTrials.gov record NCT02094209. Inclusion in this directory is not an endorsement.