Comparison of VF Induction Techniques During Medtronic ICD Implant (VF) (ICD)

Comparison of Ventricular Fibrillation Induction Techniques During Medtronic Implantable Cardioverter Defibrillator Implant

Status
Completed
Phase
N/A
Study type
Interventional
Enrollment
100 (actual)
Sponsor
James Merrill · Academic / Other
Sex
All
Age
18 Years
Healthy volunteers
Not accepted

Summary

As the indications for Implantable Cardioverter Defibrillator implantation expand, minimizing implant time is critical. Also, patients receiving biventricular ICDs are sometimes more unstable and minimization of sedation time is crucial. Multiple induction attempts, with a 1-Joule shock, can cause disruption in lead position. Therefore limiting the number of attempts will allow for better lead stability throughout the procedure and a more straightforward implant process. Investigator proposes a detailed documentation of success rates from various Ventriculart Fibrillation induction methods during implant of Medtronic defibrillation capable devices.

Detailed description

Our clinical experience has found that attempts to induce patients receiving ICDs at implant using the nominal Medtronic T-shock settings are not always effective at inducing ventricular fibrillation. When this results, a change to the parameters for T-shock is made before reattempting to induce ventricular fibrillation in patients undergoing ICD implantation. Another available induction method is 50 Hz Burst. "The 50 Hz Burst induction is designed to induce VF by delivering a rapid burst of pacing pulses. The amplitude and pulse width of these pulses are selectable, but the pacing interval is fixed at 20 ms."2 If multiple attempts using the T-shock method are unsuccessful, 50 Hz Burst provides an additional method for inducing ventricular fibrillation. It is a less desirable method since it can result in very fine VF that is difficult to identify the initiation of possibly resulting in longer times in VF causing higher defibrillation threshold or in undersensing due to the fine nature of the rhythm. T-shock has been found to be a fast and reliable method for inducing ventricular fibrillation in ICD implants. Greater success of inducing ventricular fibrillation using the T-shock method has been found when the energy of the T-shock was higher and the coupling interval was shorter. In addition, increasing the shock strength increases the window of vulnerability. For the T-shock method to be most efficient, initial success at inducing ventricular fibrillation, and therefore a reduced number of attempts, is important. This requires a better understanding of the optimal programmed settings for the initial T-shock induction attempt. Moreover, in recent years, ICD indications have grown to include primary prevention patients. These patients may have different vulnerable periods than patients previously evaluated to develop Medtronic nominal settings. Our clinical experience has found that the nominal T-shock settings are only effective at inducing ventricular fibrillation approximately seventy-five percent of the time. However, a detailed collection of success rates has not been performed.

Conditions

Interventions

TypeNameDescription
DEVICENominal T shock settingNominal Parameter Set 1 Programming Values for EnTrust
DEVICEEducated T shock settingExperimental Parameter Set 2 Programming Values

Timeline

Start date
2011-01-01
Primary completion
2012-08-01
Completion
2012-08-01
First posted
2014-01-06
Last updated
2018-05-22
Results posted
2014-12-30

Locations

1 site across 1 country: United States

Source: ClinicalTrials.gov record NCT02027883. Inclusion in this directory is not an endorsement.

Comparison of VF Induction Techniques During Medtronic ICD Implant (VF) (ICD) (NCT02027883) · Clinical Trials Directory