Closed-Loop Fluid Administration System Using Hemodynamic Monitors

Status
Completed
Phase
N/A
Study type
Interventional
Enrollment
40 (actual)
Sponsor
University of California, Irvine · Academic / Other
Sex
All
Age
18 Years
Healthy volunteers
Not accepted

Summary

The purpose of this study is to evaluate the feasibility of a closed-loop (automated) fluid administration system to deliver fluids using feedback from standard operating room hemodynamic monitors. In standard anesthesia care the rate of IV fluid administration to the patient is determined by the anesthesiologist caring for the patient. In this study protocol, the rate of fluid administration will be determined instead by the closed-loop (automated) control system under the supervision of the anesthesiologist. The system will independently decide when to give fluid and at what rate; the supervising anesthesiologist will monitor the system to ensure appropriate volumes are being delivered and to intervene if necessary.

Detailed description

In our protocol we will test the hypothesis that closed-loop fluid administration can maintain higher cardiac stroke volume (shown to improve patient outcomes after surgery) than anesthesiologist management. We have shown this to be the case in multiple simulation studies and in-vivo animal studies, but not yet experimentally in a clinical setting. Investigators will recruit their patients from the surgical record that that require cardiac output monitoring undergoing surgery and require mechanical ventilation and consent to participate in the study will be included in the study. They will receive standard patient care in that in no way will their anesthetic or surgical procedure will be altered as part of the study, with the exception of fluid administration. Fluids are usually given to a patient based on the physician's discretion. As part of the study patients will receive fluid management via a closed loop (automated) system that will use an infusion pump (Q-Core) and a controller (a computer run index and algorithm developed by Sironis) to make frequent, regular and accurate adjustments to the amount of fluid the patient receives using feedback from standard operating room monitors at UCI). Data from this study will be retrospectively compared to subjects case matched to evaluate differences in cardiac output, total fluid given, and patient outcomes.

Conditions

Interventions

TypeNameDescription
DEVICEClosed LoopFluid management in the closed loop group will be performed via a closed loop (automated) system that will use an infusion pump (Q-Core) and a controller (a computer run index and algorithm developed by Sironis) to make frequent, regular and accurate adjustments to the amount of fluid the patient receives using feedback from standard operating room monitors.

Timeline

Start date
2012-11-01
Primary completion
2014-05-01
Completion
2014-05-01
First posted
2013-12-25
Last updated
2016-05-27
Results posted
2016-05-27

Locations

1 site across 1 country: United States

Source: ClinicalTrials.gov record NCT02020863. Inclusion in this directory is not an endorsement.