Clinical Trials Directory

Trials / Completed

CompletedNCT01997502

Accuracy of Pulse CO-Oximetry Technology in Characterizing Noninvasive SpHb

Prospective Observational Study Characterizing Noninvasive Hemoglobin (SpHb) Measured With Pulse CO-Oximetry Technology in a Variety of Surgical Cases

Status
Completed
Phase
Study type
Observational
Enrollment
621 (actual)
Sponsor
University of California, Davis · Academic / Other
Sex
All
Age
18 Years
Healthy volunteers
Not accepted

Summary

This protocol is a request from Masimo to assist in the collection of data to be used to further refine the accuracy of the monitor's algorithm.

Detailed description

The specific aim of this study is to collect data to characterize Masimo's noninvasive Rainbow SpHb technology during surgeries. The purpose is to compare Masimo Rainbow technology to arterial blood samples taken during surgery. Hemoglobin concentrations in the arterial blood samples will be measured using standard blood analyzers on site and/or the sponsor's laboratory.

Conditions

Interventions

TypeNameDescription
DEVICEPulse CO-Oximetry TechnologyIn addition to the standard pulse oximeter used for clinical care, up to six SpHb finger sensors will be applied to the patient. Additional sensors may also be applied to the forehead, neck, chest and ears. Masimo FDA cleared pulse CO-Oximeter devices: Radical-7, Rad-87, Rad-57, Pronto, Pronto-7, for the noninvasive measurement of SpO2, PR, PI, PVI, SpCO, SpMet, and SpHb, total hemoglobin and respiratory rate. No treatment decisions will be made based on the new technology and no extra medical procedures are required for the study.
OTHERBlood draw4mL arterial blood samples (up to 1 per hour), outside standard of care, are drawn throughout the surgical procedure and sent to UCDMC lab/Masimo lab for analysis.

Timeline

Start date
2013-11-26
Primary completion
2024-10-21
Completion
2024-10-21
First posted
2013-11-28
Last updated
2025-04-09

Locations

1 site across 1 country: United States

Regulatory