Trials / Completed
CompletedNCT01765530
Efficacy Study of a Novel Device to Clean the Endotracheal Tube
Endotracheal Tubes Cleaned With a Novel Mechanism for Secretions Removal
- Status
- Completed
- Phase
- N/A
- Study type
- Interventional
- Enrollment
- 74 (actual)
- Sponsor
- Massachusetts General Hospital · Academic / Other
- Sex
- All
- Age
- 18 Years – 99 Years
- Healthy volunteers
- Not accepted
Summary
The purpose of this study is to evaluate the in-vivo efficacy of a novel device (endOclear) for cleaning the endotracheal tube (ETT) lumen from secretions. The investigators hypothesize that a protocol of routine ETT cleaning with endOclear may increase the ETT luminal volume measured at extubation compared to the current standard of care. The device may therefore be clinically useful by better maintaining the ETT original function.
Detailed description
After a few days of mechanical ventilation (MV), the lumen of the endotracheal tube (ETT) is coated with a thick bacterial biofilm, which is a potential source for bacterial colonization of the lower respiratory tract and ventilator-associated pneumonia (VAP). Accumulation of mucus/secretions on the interior of the ETT also effectively lowers the cross section of the ETT and increases significantly the work of breathing in intubated patients, who then require increased MV support, with prolonged intubation and ICU stay. A primary goal for all ICUs is to shorten the time of MV, to decrease the incidence of ETT microbial colonization and ultimately VAP incidence. To meet this goal several airways technologies to maintain the ETT free from debris/bacteria have been developed in order to improve the current ventilator-liberation standards in intubated critically ill patients. Based on preliminary laboratory, animal and clinical data, The investigators now wish to test the hypothesis that a novel commercially available ETT cleaning device can be effective in targeting two specific issues: * Specific aim 1: To evaluate the mean difference in ETT luminal volume of ETTs cleaned with the novel cleaning device (EndOclear) vs. ETT cleaned in the standard manner. * Specific aim 2: To determine whether ETTs cleaned with the novel device show a lower degree of microbial colonization for the entire period of mechanical ventilation. Data about in-vivo respiratory mechanics measurements, microbial diversity and antibiotic resistance analysis, users' feedbacks and safety of the device will also be collected. The investigators will conduct a clinical trial on 74 intubated patients randomized to receive standard ETT cleaning or cleaning with the EndOclear.
Conditions
- Endotracheal Extubation
- Airway Obstruction
- Airway Control
- Pneumonia, Ventilator-Associated
- Breathing Mechanics
Interventions
| Type | Name | Description |
|---|---|---|
| DEVICE | ETT cleaning maneuver | EndOclear device will be used for ETT cleaning maneuver. This is a commercially available device. It consists of an ETT cleaning apparatus with a flexible central tube and a cleaning device at its distal end. The cleaning device has a mechanically expandable mesh structure which can be adapted to selectively move between a radially-collapsed position and a radially-expanded one. In operation, the collapsed cleaning apparatus is inserted into the ETT through a Y-shaped connector. The device is then expanded by a safety toggle protected trigger mechanism that, when fired, presses the device's smooth silicone disc against the inside surface of the ETT. The cleaning apparatus is then pulled out of the ETT removing mucus deposits and secretions. We will add endOclear to the standard ICU practice, scheduling the systematic use of the device every 8 hours for the whole intubation period. |
Timeline
- Start date
- 2012-09-01
- Primary completion
- 2013-12-01
- Completion
- 2013-12-01
- First posted
- 2013-01-10
- Last updated
- 2017-10-18
- Results posted
- 2017-03-29
Locations
1 site across 1 country: United States
Source: ClinicalTrials.gov record NCT01765530. Inclusion in this directory is not an endorsement.