CompletedNCT01714232

Evaluation of Blood Glucose Meter Systems

Evaluation of Blood Glucose Meter Systems - Contour® PLUS Study

Status
Completed
Phase
N/A
Study type
Interventional
Enrollment
106 (actual)
Sponsor
Ascensia Diabetes Care · Industry
Sex
All
Age
18 Years
Healthy volunteers
Accepted

Summary

The purpose of this study was to evaluate the performance of a Bayer Blood Glucose Monitoring System (BGMS) and four additional Blood Glucose Monitoring Systems (BGMS) from other manufacturers. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).

Detailed description

The purpose of this study was to evaluate the performance of a Bayer Blood Glucose Monitoring System (BGMS) and four additional Blood Glucose Monitoring Systems (BGMS) from other manufacturers. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer). Performance of the five systems were evaluated across the glucose range of the BGMSs using capillary blood. All testing and lancing were performed by study staff and some samples were tested from subject fingertips. Additionally, some blood samples were glycolyzed to lower the glucose concentration levels and glucose solution was added to other samples to raise glucose concentration levels.

Conditions

Interventions

TypeNameDescription
DEVICEContour® PLUS BGMSStudy staff performed Blood Glucose (BG) tests with capillary fingerstick blood collected from subjects with diabetes and without diabetes (up to 10% of subjects without diabetes were included). All meter BG results were compared with capillary plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).
DEVICEOneTouch® SelectSimple™ BGMSStudy staff performed Blood Glucose (BG) tests with capillary fingerstick blood collected from subjects with diabetes and without diabetes (up to 10% of subjects without diabetes were included). All meter BG results were compared with capillary plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).
DEVICEAccu-Chek® Performa BGMSStudy staff performed Blood Glucose (BG) tests with capillary fingerstick blood collected from subjects with diabetes and without diabetes (up to 10% of subjects without diabetes were included). All meter BG results were compared with capillary plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).
DEVICEAccu-Chek® Active BGMSStudy staff performed Blood Glucose (BG) tests with capillary fingerstick blood collected from subjects with diabetes and without diabetes (up to 10% of subjects without diabetes were included). All meter BG results were compared with capillary plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).
DEVICEFreestyle Freedom® BGMSStudy staff performed Blood Glucose (BG) tests with capillary fingerstick blood collected from subjects with diabetes and without diabetes (up to 10% of subjects without diabetes were included). All meter BG results were compared with capillary plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).

Timeline

Start date
2012-10-01
Primary completion
2012-11-01
Completion
2012-11-01
First posted
2012-10-25
Last updated
2016-02-29
Results posted
2013-12-27

Locations

1 site across 1 country: United States

Source: ClinicalTrials.gov record NCT01714232. Inclusion in this directory is not an endorsement.