Clinical Trials Directory

Trials / Completed

CompletedNCT01636804

Incidence & Predictors of Mechanical Abnormalities of QuickSite® and QuickFlex® Left Ventricular Pacing Leads

Incidence & Predictors of Mechanical Abnormalities of QuickSite® and QuickFlex® Left Ventricular Pacing Leads: Impact on Electrophysiological Profile

Status
Completed
Phase
Study type
Observational
Enrollment
190 (estimated)
Sponsor
Dhanunjaya Lakkireddy, MD, FACC · Academic / Other
Sex
All
Age
18 Years
Healthy volunteers
Not accepted

Summary

On April 2, 2012 the manufacturer of the Quicksite® and Quickflex® leads, St. Jude Medical, released a voluntary medical device advisory regarding these leads' potential for problems with the silicone insulation covering one end of the lead, where it meets another section of the lead which is coated with a different insulating material (polyurethane). There has been a voluntary suspension of sales of these leads. There are approximately 179,000 of the above leads implanted worldwide with an estimated 101,000 leads currently in service (59,000 in US). There have been 39 confirmed cases of the conductor portion of the lead becoming exposed from the silicone insulation (externalization), and the rate of occurrence of this type of problem is conservatively estimated at less than one percent. However, a review done by St. Jude Medical of returned Quicksite® and Quickflex® leads and available X-ray and fluoroscopic (x-ray movie) images of patients with these leads (from 1,219 patients) revealed a higher rate of 3-4%. It is presumed that there are patients whose leads have not yet been affected by this problem, but whose leads may show signs of structural changes on x-ray or fluoroscopy. This study seeks to determine what structural changes may be occurring in these leads, what signs might predict these changes, and what impact these changes may have on the lead and its performance.

Conditions

Interventions

TypeNameDescription
DEVICEQuicksiteSt. Jude Medical produced lead called Quicksite
DEVICEQuickflexSt. Jude Medical produced lead called Quickflex

Timeline

Start date
2012-07-01
First posted
2012-07-10
Last updated
2013-08-09

Locations

1 site across 1 country: United States

Source: ClinicalTrials.gov record NCT01636804. Inclusion in this directory is not an endorsement.