Trials / Completed
CompletedNCT01322737
Feasibility Trial to Evaluate the Ability of the SuMO Tissue Access and Resection System
Evaluation of a Novel Endoscopic Polypectomy Device in Excised Human Colon and Gastric Tissue.
- Status
- Completed
- Phase
- N/A
- Study type
- Interventional
- Enrollment
- 9 (actual)
- Sponsor
- University Hospitals Cleveland Medical Center · Academic / Other
- Sex
- All
- Age
- 18 Years
- Healthy volunteers
- Accepted
Summary
This feasibility trial will evaluate the ability of the SuMO Tissue Access and Resection System (Apollo Endosurgery, Austin, Texas). The SuMO System utilizes balloons to create a submucosal pocket and electrosurgical cutting device to resect the mucosa. The SuMO elevates the tissue from the underlying muscularis using a balloon to distend the submucosa, similar to what is currently performed with injectable saline in a standard EMR. After dissection of the lesion, a separate cutting device will be utilized to resect the tissue by cutting around the border of the targeted tissue. If needed, a standard loop snare will be used to complete the mucosal resection.
Detailed description
The subject will already be scheduled for segmental colectomy or gastrectomy. The colectomy or gastrectomy specimen will be prepared on a side table for an ex vivo endoscopic procedure. The ends of the stomach or colon, if not already stapled closed will be ligated to allow for insufflation. The endoscope will be placed into the excised organ and the SuMO System devices will be deployed away from any areas of excised pathologic tissue, so as not to disrupt the pathologist's evaluation of the excised tissue. The specimen will then be opened and grossly examined for the actual completeness of the submucosal pocket and resection. Mucosal and muscular wall disruptions will be assessed. Again, no interference in the underlying pathology will be performed. If sufficient specimen is present, the SuMO procedure may be repeated on the same specimen. Following the performance of the EMR by the SUMO balloon system on the ex vivo colon specimen, both the resected piece of tissue and the site of resection will be evaluated pathologically by H and E staining. Specifically, the depth of resection in the ex vivo colon specimen will be determined. In addition, the level of dissection on the removed piece of mucosa will also be identified. During this procedure still images or a video recording maybe captured, however all images will be taken of the removed specimen. There will not be any patient identifiers on any of the images or DVD's taken during the procedure. No clinical follow up is required, as the treated area will have already been resected and there is no risk to the patient.
Conditions
Interventions
| Type | Name | Description |
|---|---|---|
| PROCEDURE | Tissue Access and Resection System | This feasibility trial will assess the effectiveness of SuMO System to create a submucosal pocket and resect the overlying mucosa in excised human colonic or gastric tissue |
Timeline
- Start date
- 2011-01-01
- Primary completion
- 2011-06-01
- Completion
- 2011-06-01
- First posted
- 2011-03-25
- Last updated
- 2014-12-16
Locations
1 site across 1 country: United States
Source: ClinicalTrials.gov record NCT01322737. Inclusion in this directory is not an endorsement.