CompletedNCT00636077

Comparative Effects Of Dialysate Flow Rate And Membrane Packing On The Performance Of Dialyzers Used For Hemodialysis

Effect Of Dialysate Flow Rate On Mass Transfer Coefficient - Area (KoA) In Dialyzers With Different Membrane Packing Densities

Status
Completed
Phase
N/A
Study type
Interventional
Enrollment
12 (actual)
Sponsor
Vantive Health LLC · Industry
Sex
All
Age
19 Years
Healthy volunteers
Not accepted

Summary

The purposes of this study are to determine if the performance of a dialyzer depends on how tightly the hollow fiber membranes are packed in the housing of the dialyzer (the membrane packing density) and if that dependence is a function of the dialysate flow rate. The study will examine how efficiently three different sized molecules pass through a dialyzer membrane at different dialysate flow rates in dialyzers with different membrane packing densities. Transfer of urea, phosphorus and beta-2-microglobulin from blood to dialysate will be measured during clinical hemodialysis using four different dialyzers, each used at three different dialysate flow rates. The data derived from these measurements may provide insight into the importance of membrane packing density as a design parameter for hemodialyzers and if changing the membrane packing density might provide equivalent performance at a lower dialysate flow rate.

Conditions

Interventions

TypeNameDescription
DEVICEPolyflux HD-C4 BigThree consecutive treatments with the HD-C4 Big. During the third treatment, dialyzer clearances of urea, phosphorus and β2-microglobulin will be determined at a blood flow rate of 400 mL/min and dialysate flow rates of 350 mL/min, 500 mL/min and 800 mL/min.
DEVICEHD-C4 SmallThree consecutive treatments with the HD-C4 Small. During the third treatment dialyzer clearances of urea, phosphorus and β2-microglobulin will be determined at a blood flow rate of 400 mL/min and dialysate flow rates of 350 mL/min, 500 mL/min and 800 mL/min.
DEVICEOptiflux 160NRThree consecutive treatments with the HD-C4 Small. During the third treatment dialyzer clearances of urea, phosphorus and β2-microglobulin will be determined at a blood flow rate of 400 mL/min and dialysate flow rates of 350 mL/min, 500 mL/min and 800 mL/min.
DEVICEOptiflux 200NRThree consecutive treatments with the HD-C4 Small. During the third treatment dialyzer clearances of urea, phosphorus and β2-microglobulin will be determined at a blood flow rate of 400 mL/min and dialysate flow rates of 350 mL/min, 500 mL/min and 800 mL/min.

Timeline

Start date
2008-02-01
Primary completion
2008-04-01
Completion
2008-04-01
First posted
2008-03-14
Last updated
2025-07-14
Results posted
2011-07-01