Clinical Trials Directory

Trials / Completed

CompletedNCT00556543

Clinical Study of the U-Plate Fracture Repair System to Treat Rib Fractures

Clinical Evaluation of the U-plate Fracture Repair System for the Fixation of Rib Fractures

Status
Completed
Phase
N/A
Study type
Interventional
Enrollment
10 (actual)
Sponsor
Oregon Health and Science University · Academic / Other
Sex
All
Age
18 Years
Healthy volunteers
Not accepted

Summary

The purpose of this study is to gather information about a device used to help fix broken ribs. Hypothesis: Rib fracture repair with the U-plate system is clinically durable and safe for the indications of flail chest repair, acute pain control, chest wall defect repair, and rib fracture non-union.

Detailed description

Rib fractures are a painful and disabling injury commonly found among trauma patients. According to the National Center for Health Statistics, approximately 300,000 people with rib fractures were treated in emergency and ambulatory care departments in the United States in the year 2004. Rib fracture patients are significantly more disabled at 30 days post-injury than patients with chronic medical illness and lose an average of 70 days of work or usual activity during their acute recovery. The rationale for conducting this study is to demonstrate in a prospective study that the U-plate repair system is durable and safe. Although this clinical outcome data is not required by the FDA to market and implant this prosthesis, the investigators believe that in order for the U-plate to be widely accepted, prospectively collected outcome data are necessary. With the goal of improving the durability of fixation of rib fractures over the techniques currently available and with the additional goal of developing a minimally invasive technique, a U-shaped plate was developed. The design of the U-plate theoretically overcomes the inherent softness of the human rib by grasping the rib over its superior margin and by securing the plate with anterior to posterior locking screws that do not rely on screw purchase in bone. Thus much of the strength and durability of the fixation is transferred from the relatively soft rib to the plate itself.

Conditions

Interventions

TypeNameDescription
DEVICEU-plate fracture repair systemAll subjects meeting inclusion criteria will undergo surgery for the repair of their fractured rib/s using the U-plate device. Subjects will be monitored daily during the entire course of their hospitalization for adverse experiences. Subjects will be contacted by phone to complete the MPQ and SF-36 Health Survey. Subjects will be asked if and when they have returned to work or to their previous level of functioning. Subjects will be asked if they had any complications or problems associated with their surgery.

Timeline

Start date
2006-11-01
Primary completion
2008-10-01
Completion
2008-10-01
First posted
2007-11-12
Last updated
2011-07-01
Results posted
2011-05-10

Locations

1 site across 1 country: United States

Source: ClinicalTrials.gov record NCT00556543. Inclusion in this directory is not an endorsement.